Transport and storage
multi EA 5000
174
}
Disconnect the measuring cell from the exhaust hose and remove the measuring cell.
Empty the measuring cell.
}
Disconnect the gas transfer line from the sulfuric acid container.
}
Remove the sulfuric acid container from the detection module. Empty and rinse the
container Replacing the sulfuric acid and cleaning the sulfuric acid container..
}
Put all electrodes into their original packaging. Observe the instructions on mainte-
nance and care of the electrodes (→ "Maintenance and storage of the electrodes"
153).
}
Clean the measuring cell with distilled water and ethanol. Carefully wipe out/off the
measuring cell and the magnetic stirring rod with tissue.
}
Fasten the exhaust hose in the detection module, e.g., with adhesive strips.
15.5.2 Notes on transport of the Coulometric Sulfur Module
}
Disconnect the electrode cables from the "Generation" and "Indication" connections.
}
Disconnect the hose connections to the HX and NOx absorbers and remove both ab-
sorbers from the module.
}
Remove the measuring cell from the module.
}
Remove the electrodes and the gas inlet tube from the measuring cell.
}
Empty the measuring cell, remove the magnetic stirring rod and rinse both with ul-
trapure water.
}
Package the glass components and the electrodes in their original packaging.
15.5.3 Notes on transport of the TOC module
WARNING
Risk of chemical burns
The TIC reactor can contain 40 % phosphoric acid residue. Phosphoric acid can irritate
eyes, skin and mucous membranes.
¡
Wear protective clothing when emptying and cleaning the TIC reactor.
¡
Observe all notes and specifications in the safety data sheet.
}
Remove the halogen trap and the water traps from the detection module.
}
Remove the TIC reactor from the detection module and rinse it out.
}
In the basic module: Remove the condensation coil and the TOC combustion tube af-
ter the furnace has cooled.
CAUTION! Risk of burns around the hot furnace!
}
Package all components and connection hoses in their original packaging.