PR-000 985_J 

 

 

 

              3/43 

Endosense TactiCath

®

  

Diagnostic/Ablation Deflectable Tip Catheter 

 

CAUTION 

••••

  Sterile. Sterilized with ethylene oxide gas. 

••••

  For single use only. 

••••

  Do not use if the package is open or damaged. 

••••

  Do  not  attempt  to  operate  the  device  prior  to  completely  reading 

and understanding the applicable instructions for use. 

CATHETER DESCRIPTION 

The  Endosense  TactiCath

®

  Diagnostic/Ablation  Deflectable  Tip  Catheter 

designed  for  cardiology  application  is  an  8.5  F  introducer  compatible 
multi-electrode luminal catheter with a deflectable tip designed to facilitate 
electrophysiological  mapping  of  the  heart  and  to  transmit  radiofrequency 
(RF) current to the catheter tip electrode for intracardial ablation purposes. 
For  ablation  the  catheter  is  used  in  conjunction  with  a  radiofrequency 
generator and a dispersive pad (indifferent patch electrode).  
 
The  catheter  has  a  high-torque  shaft  with  a  deflectable  tip  section  with 
four  platinum-iridium  electrodes.  All  of  the  electrodes  may  be  used  for 
recording  and  stimulation  purposes.  The  tip  electrode  is  used  to  deliver 
radiofrequency  current  from  the  radiofrequency  generator  to  the  desired 
ablation site. The tip electrode and ring electrodes are made of platinum-
iridium.  The  catheter  incorporates  a  thermocouple  temperature  sensor 
which is embedded in the 3.5 mm tip electrode. Tip deflection is controlled 
at the proximal end by a handpiece in which a piston slides; a thumbknob 
on  the  piston  controls  piston  travel.  When  the  thumbknob  is  pushed 
forward the tip  is deflected (curved). When the thumbknob is pulled back 
the tip straightens. The shape of the curve depends on the deflectable tip 
length (currently available: 65 mm). The high-torque shaft also allows the 
plane of the curved tip to be rotated to facilitate accurate positioning of the 
catheter tip at the desired site. 
 
At the proximal end of the catheter lead-out a saline port with a standard 
Luer  fitting  terminates  from  the  open  lumen.  This  saline  port  is  used  to 
permit  the  delivery  of  isotonic  saline  solution  to  irrigate  the  tip  electrode. 
During  ablation  normal  saline  is  passed  through  the  0.7  mm  diameter 
lumen  of  the  catheter  and  through  the  tip  electrode,  to  irrigate  and  cool 
both the catheter tip and the ablation site. 
 
The  catheter  interfaces  with  standard  recording  equipment  and  a 
compatible  radiofrequency  generator  via  accessory  Endosense  Splitter 
with the appropriate connectors. 
 
The distal force sensing technology called Touch+

®

 is also an integral part 

of  Endosense  TactiCath

®

  Diagnostic/Ablation  Deflectable  Tip  Catheter. 

For use in conjunction with force  sensing refer  to  the TactiCath

®

 System 

User Manual. 

INDICATIONS 

The  Endosense  TactiCath

®

  Diagnostic/Ablation  Deflectable  Tip  Catheter 

is  indicated  for  use  in  cardiac  electrophysiological  mapping  (stimulation 
and  recording)  and,  when  used  in  conjunction  with  a  radiofrequency 
generator,  for  cardiac  ablation  in  the  right  atrium  (RA)  and  in  the  left 
atrium (LA), such as: 

••••

 

Interruption  of  accessory  atrioventricular  (AV)  conduction 
pathways  associated  with  tachycardia,  called  the  Wolff-
Parkinson-White syndrome (WPW); 

••••

 

Treatment of AV nodal re-entrant tachycardia (AVNRT);  

••••

 

Creation of complete AV nodal block in patients with a difficult to 
control ventricular response to an atrial arrhythmia; 

••••

 

Creation  of  a  RA  isthmus  line  for  patients  with  isthmus 
dependent atrial flutter; 

••••

 

Atrial fibrillation. 

CONTRAINDICATIONS 

••••

  The catheter must not be used for radiofrequency ablation if the patient 

has  intracardiac  mural  thrombus  or  has  had  a  ventriculotomy  or 
atriotomy within the preceding four weeks.  

••••

  The catheter may not be appropriate for patients with prosthetic valves 

and must not be used in the coronary vasculature. 

••••

  A  relative  contraindication  for  catheter  ablation  procedures  is  active 

systemic infection. 

••••

  The  transseptal  approach  is  contraindicated  in  patients  with  left  atrial 

thrombus or myxoma, or interatrial baffle or patch. 

WARNINGS 

a) 

The  temperature  sensor  located  within  the  electrode  will  not  reflect 
either  electrode  tissue  interface  or  tissue  temperature  due  to  the 
cooling  effects  of  the  saline  irrigation  of  the  electrode.  The 
temperature  displayed  on  the  generator  is  the  temperature  of  the 
cooled electrode, not tissue temperature. The temperature sensor is 
used  to  verify  that  the  irrigation  flow  rate  is  adequate.  Before 
initiating  the  application  of  radiofrequency  current,  a  decrease  in 
electrode  temperature  confirms  the  onset  of  saline  irrigation  of  the 
ablation electrode. Recording temperature from the electrode during 
the  application  of  radiofrequency  current  ensures  that  the  irrigation 
flow rate is being maintained. 

b) 

Patients undergoing septal accessory pathway ablation are at risk for 
complete  AV  block  which  requires  the  implantation  of  a  permanent 
pacemaker.  Permanent  pacing  may  be  required  in  patients  who 
experience  inadvertent  complete  AV  block  as  a  result  of 
radiofrequency ablation. 

c) 

Implantable  pacemakers  and  implantable  cardioverter/defibrillator 
(ICDs)  may  be  adversely  affected  by  radiofrequency  current.  It  is 
important to: 

••••

 

Have  temporary  external  sources  of  pacing  and 
defibrillation available during ablation. 

••••

 

Temporarily  reprogram  the  pacing  system  to  minimum 
output to minimize risk of inappropriate pacing. 

••••

 

Exercise  extreme  caution  during  ablation  when  in  close 
proximity to atrial or ventricular permanent leads. 

••••

 

Program  the  ICD  to  the  OFF  mode  during  the  ablation 
procedure. 

••••

 

Perform  complete  implantable  device  analysis  on  all 
patients after ablation. 

 

d) 

The combination of  intracoronary placement of the ablation catheter 
and  radiofrequency  energy  application  has  been  associated  with 
myocardial infarction. 

e) 

Significant x-ray exposure can result in acute radiation injury as well 
as  increased  risk  for  somatic  and  genetic  effects,  to  both  patients 
and  laboratory  staff.  Therefore  catheter  ablation  should  only  be 
performed  after  adequate  attention  has  been  given  to  the  potential 
radiation exposure associated with the procedure, and steps taken to 
minimize  this  exposure.  Careful  consideration  must  therefore  be 
given for the use of the device in pregnant women. 

f) 

Do not expose catheter to organic solvents such as alcohol. 

g) 

Do not resterilize/reuse the catheter. 

h) 

Do  not  immerse  proximal  handle  or  cable  connector  in  fluids; 
electrical performance could be affected. 

i) 

Do not scrub or twist the distal tip electrode during cleaning. 

j) 

Inspect  saline  irrigation  for  air  bubbles  prior  to  its  use  in  the 
procedure. Air bubbles in the saline irrigation may cause emboli. 

PRECAUTIONS 

a) 

Do 

not 

attempt 

to 

operate 

the 

Endosense 

TactiCath

®

 

Diagnostic/Ablation  Deflectable  Tip  Catheter  or  the  radiofrequency 
generator  prior  to  completely  reading  and  understanding  the 
applicable instructions for use. 

b) 

Cardiac  ablation  procedures  must  be  performed  by  appropriately 
trained personnel in a fully-equipped electrophysiology laboratory. 

c) 

The  long-term  risks  of  protracted  fluoroscopy  and  creation  of 
radiofrequency  induced  lesions  have  not  been  established.  Careful 
consideration  must  therefore  be  given  for  the  use  of  the  device  in 
prepubescent children. Furthermore, the risk/benefit in asymptomatic 
patients has not been studied. 

Содержание TactiCath

Страница 1: ...unta direzionabile Italiano Pagina 15 Instrucciones de uso Endosense TactiCath Cat ter para diagn stico ablaci n con punta dirigible Espa ol P gina 19 Instru es de utiliza o Endosense TactiCath Catete...

Страница 2: ......

Страница 3: ...ng Pouze k jednor zov mu pou it Dal informace najdete v n vodu k pou it Use by Serial number Utiliser avant Num ro de s rie Verwendbar bis Seriennummer Usare entro Numero di serie Fecha de caducidad N...

Страница 4: ...diac mural thrombus or has had a ventriculotomy or atriotomy within the preceding four weeks The catheter may not be appropriate for patients with prosthetic valves and must not be used in the coronar...

Страница 5: ...nerator during normal operation may adversely affect the performance of other equipment u Electrodes and probes for monitoring and stimulating devices can provide paths for high frequency current The...

Страница 6: ...ion flow 9 5 Dealing with irrigation hole occlusion If irrigation hole occlusion occurs proceed as follows 1 Remove the catheter from the patient 2 Fill a syringe with sterile saline and attach to the...

Страница 7: ...ARISING OUT OF THE REUSE OF ANY PRODUCT S LABELED FOR SINGLE USE OR WHERE REUSE IS PROHIBITED BY APPLICABLE LAW AS WELL AS ANY REPROCESSING OR RESTERILIZATION OF THE PRODUCT S DESCRIBED HEREIN Handlin...

Страница 8: ...rome de Wolff Parkinson White WPW Traitement d une tachycardie par r entr e intra nodale AV Cr ation d un bloc complet du n ud AV pour les patients chez qui la r ponse ventriculaire une arythmie atria...

Страница 9: ...ence de d fauts vidents ou une mauvaise application k Lire et suivre le mode d emploi fourni par le fabricant de l lectrode indiff rente l utilisation d lectrodes indiff rentes satisfaisant ou d passa...

Страница 10: ...nutes pour obtenir la stabilisation de la mesure de la force dans le cath ter Si la d rive de la force d passe 5 grammes en l espace de 2 minutes laisser le cath ter dans la veine cave pendant 2 minut...

Страница 11: ...UI A ONT T FABRIQU S ET LIB R S POUR UNE UTILISATION CONFORME SES LEURS SP CIFICATIONS ET QUI ONT T TEST S AVEC DES M THODES DE TESTS RECONNUES ENDOSENSE LTD NE DONNE AUCUNE AUTRE GARANTIE EXPR S OU I...

Страница 12: ...atienten mit gest rter ventrikul rer Antwort auf eine atriale Arrhythmie Erzeugung einer RV Isthmus Linie f r Patienten mit Isthmus abh ngigem Vorhofflattern Vorhofflimmern 3 GEGENANZEIGEN Der Kathete...

Страница 13: ...nten Elektrode die Verwendung von indifferenten Elektroden die mindestens den ANSI AAMI Anforderungen entsprechen HF18 empfiehlt sich l Der steuerbare Diagnostik Ablationskatheter Endosense TactiCath...

Страница 14: ...Katheter 2 Minuten lang im Blut des Patienten bevor Sie die volle Leistung erwarten Das Eintauchen gestattet eine Stabilisierung der Kraftmessung im Katheter Wenn nach 2 Minuten die Kraftabweichung b...

Страница 15: ...METHODEN GETESTET WURDE ENDOSENSE LTD GEW HRT NEBEN DER HIER GENANNTEN KEINE AUSDR CKLICHE ODER STILLSCHWEIGENDE GARANTIE UND SCHLIE T EINE GARANTIE DER MARKTF HIGKEIT ODER DER EIGNUNG DES DER HIER BE...

Страница 16: ...aritmia atriale Esecuzione di linea istmo AD per pazienti con flutter atriale istmo dipendente Fibrillazione Atriale 3 CONTROINDICAZIONI Il catetere non deve essere usato per l ablazione con radiofreq...

Страница 17: ...so p Non risterilizzare n riutilizzare q Il Catetere diagnostico per ablazione con punta direzionabile Endosense TactiCath usato insieme a un generatore di radiofrequenza in grado di erogare una signi...

Страница 18: ...essersi assicurati che l elettrodo in punta sia in contatto stabile con il sito dell ablazione avviare l erogazione di corrente di radiofrequenza L impedenza del circuito deve essere approssimativamen...

Страница 19: ...SUE SOCIET AFFILIATE VERRANNO RITENUTE RESPONSABILI DI QUALSIASI EVENTUALE PERDITA DANNO O COSTO SPECIALE DIRETTO INDIRETTO INCIDENTALE O EMERGENTE O DI ALTRE PERDITE DANNI O COSTI DIRETTI O INDIRETTI...

Страница 20: ...icular a la arritmia atrial Creaci n de una l nea de istmo del AD para pacientes con aleteo atrial dependiente del istmo Fibrilaci n atrial 3 CONTRAINDICACIONES El cat ter no debe ser utilizado para a...

Страница 21: ...cat ter para diagn stico ablaci n con punta dirigible TactiCath de Endosense ha sido dise ado para un solo uso p No volver a esterilizar ni reusar q El Cat ter para diagn stico ablaci n con punta diri...

Страница 22: ...uado 3 Utilizar la perilla para facilitar la colocaci n de la punta del cat ter Si se desea que la punta del cat ter Entonces la perilla se desv e empuje se enderece tire hacia atr s Precauci n No se...

Страница 23: ...O S HA N SIDO FABRICADO S Y LIBERADO S PARA SU USO DE ACUERDO CON SUS ESPECIFICACIONES Y PROBADOS UTILIZANDO M TODOS ESTABLECIDOS DE PRUEBA ENDOSENSE LTD NO OTORGA NINGUNA OTRA GARANT A EXPRESA O IMPL...

Страница 24: ...odal AV TRNAV Cria o de um bloqueio total do n dulo AV em doentes com dificuldade em controlar a resposta ventricular a arritmias auriculares Cria o de uma linha do istmo na AD para doentes com flutte...

Страница 25: ...o Endosense TactiCath destina se a uma nica utiliza o p N o voltar a esterilizar e reutilizar q O cateter de ponta dirig vel para diagn stico abla o Endosense TactiCath utilizado em conjunto com o ger...

Страница 26: ...bot o se deflicta empurre para a frente se endireite puxe para tr s Aten o N o deve utilizar liqu do de contraste no cateter 9 3 Aplica o de corrente de radiofrequ ncia 1 Estabelecer um caudal de irri...

Страница 27: ...EM QUE O S PRODUTO S FOI FORAM FABRICADOS E INTRODUZIDOS NO MERCADO PARA UTILIZA O DE ACORDO COM AS SUAS ESPECIFICA ES E TESTADOS PELOS SEUS M TODOS DE TESTE ESTABELECIDOS A ENDOSENSE LTD N O PROPORC...

Страница 28: ...Tot stand brengen van een RA isthmus lijn voor pati nten met isthmus afhankelijke atriumflutter Atriumfibrilleren 3 CONTRA INDICATIES De katheter mag niet worden gebruikt voor radiofrequentie ablatie...

Страница 29: ...aats vindt in de buurt van structuren zoals de sinusknoop en de AV knoop n V r gebruik moet de steriele verpakking en de katheter worden gecontroleerd o De Endosense TactiCath Diagnose Ablatie Kathete...

Страница 30: ...om in aan de hoogste stroomsnelheid zoals weergegeven in bovenstaande tabel Kijk hoe de temperatuur van de elektrodepunt afneemt 2 Wanneer besloten is dat de puntelektrode een stabiel contact heeft me...

Страница 31: ...RDT ENDOSENSE LTD OF ZIJN FILIALEN ENIGE AANSPRAKELIJKHEID VOOR SPECIALE DIRECTE INDIRECTE INCIDENTELE EN GEVOLGSCHADE OF ANDERE SCHADE OF KOSTEN DIE DIRECT DAN WEL INDIRECT VOORTVLOEIEN UIT HET GEBRU...

Страница 32: ...is patienten har intrakardielle fastsiddende tromber og har gennemg et et ventrikulotomi eller atriotomi inden for de sidste fire uger Katetret er muligvis ikke bedste valg til patienter med kunstige...

Страница 33: ...Cath diagnostisk ablationskatetret med fleksibel spids n r det anvendes sammen med en RF generator under normale forhold kan have negativ indvirkning p ydeevnen af andet udstyr u Elektroder og sonder...

Страница 34: ...d tilstopning af skyllehuller Hvis skyllehullerne tilstoppes iv rks ttes f lgende tiltag 1 Fjern katetret fra patienten 2 Fyld en spr jte med sterilt saltvand og tilslut den til systemet ideelt ved hj...

Страница 35: ...NBRUG AF ET PRODUKT DER ER M RKET KUN TIL ENGANGSBRUG ELLER HVOR GENBRUG IKKE ER TILLADT IF LGE G LDENDE LOV SAMT EVENTUEL OMBEARBEJDNING ELLER RESTERILISATION AF DE HERI BESKREVNE PRODUKTER H ndterin...

Страница 36: ...RAINDIKATIONER Katetern f r inte anv ndas f r radiofrekvensablation om patienten har intrakardiell mural tromb eller har genomg tt en ventrikulotomi eller atriotomi inom de f reg ende fyra veckorna Ka...

Страница 37: ...trokirurgi s F rsiktighets tg rder m ste tas f r att avgr nsa l ttant ndligt materialer fr n det elektrokirurgiska omr det t Elektromagnetisk st rning EMI som skapas av Endosense TactiCath Diagnostik...

Страница 38: ...kluderade genom att avlufta dem med ett spolningsfl de med h g hastighet innan terinf randet 9 5 F rfarande vid ocklusion i spolningsh l Forts tt enligt f ljande om ocklusion i spolningsh l uppst r 1...

Страница 39: ...SKADOR SOM UPPST R GENOM TERANV NDNING AV N GON PRODUKT SOM ANGES F R ENG NGSBRUK ELLER D R TERANV NDNING R F RBJUDEN I TILL MPLIG LAG LIKSOM N GON UPPARBETNING ELLER OMSTERILISERING AV PRODUKTER SOM...

Страница 40: ...t ln cesta je kontraindikov na u pacient s trombem nebo myxomem v lev s ni nebo mezis ov m uz v rem i z platou 4 VAROV N a Tepeln idlo v elektrod vzhledem k jej mu chlazen iriga n m roztokem nem teplo...

Страница 41: ...detek n m syst mu kterou je p ed aplikac radiofrekven n energie nutno vy e it w P ed pou it m syst mu zkontrolujte zda jsou iriga n porty pln funk n a to aplikac fyziologick ho roztoku katetrem x Pra...

Страница 42: ...acienta Varov n Nepou vejte d le katetr pokud je st le ucpan nebo pokud nefunguje spr vn Pozn mka Mal st ka ka zajist dostate n tlak k vytvo en viditeln ho proudu kapaliny 10 NE DOUC REAKCE P i katetr...

Страница 43: ...ODY NEBO N KLADY VZNIKL OPAKOVAN M POU IT M JAK HOKOLI V ROBKU OZNA EN HO JAKO JEDNOR ZOV NEBO KDE JE OPAKOVAN POU IT ZAK Z NO P SLU N M Z KONEM A TAK P I JAK KOLI STERILIZACI NEBO P PRAV ZDE POPSAN H...

Страница 44: ...du Grand Puits 42 1217 Meyrin Switzerland Tel 41 22 306 19 58 Fax 41 22 306 19 78 www endosense com PATENTS AND TRADEMARKS Patents pending Endosense the Endosense logo TactiCath Touch are registered t...

Отзывы: