59
Specifications and Reference Data
Specifications and Reference Data
July 2016
Reference Manual
00809-0200-4648, Rev BB
A.1.4 Accuracy
Table A-2. Rosemount 648 Wireless Input Options and Accuracy
Sensor options
Sensor reference
Input ranges
Digital accuracy
(1)
2-, 3-, 4-wire RTDs
°C
°F
°C
°F
Pt 100 (
= 0.00385)
IEC 751
–200 to 850
–328 to 1562
± 0.225
± 0.405
Rosemount X-well Pt
100 (0.00385)
IEC 751
–50 to 300
–58 to 572
± 0.29
± 0.52
Pt 200 (
= 0.00385)
IEC 751
–200 to 850
–328 to 1562
± 0.405
± 0.729
Pt 500 (
= 0.00385)
IEC 751
–200 to 850
–328 to 1562
± 0.285
± 0.513
Pt 1000 (
= 0.00385)
IEC 751
–200 to 300
–328 to 572
± 0.285
± 0.513
Pt 100 (
= 0.003916)
JIS 1604
–200 to 645
–328 to 1193
± 0.225
± 0.405
Pt 200 (
= 0.003916)
JIS 1604
–200 to 645
–328 to 1193
± 0.405
± 0.729
Ni 120
Edison Curve No. 7
–70 to 300
–94 to 572
± 0.225
± 0.405
Cu 10
Edison Copper Winding No. 15
–50 to 250
–58 to 482
± 2.1
± 3.78
Pt 50 (
= 0.00391)
GOST 6651-94
–200 to 550
–328 to 990
± 0.45
± 0.81
Pt 100 (
= 0.00391)
GOST 6651-94
–200 to 550
–328 to 990
± 0.225
± 0.405
Cu 50 (
= 0.00426)
GOST 6651-94
–50 to 200
–58 to 392
± 0.72
± 1.296
Cu 50 (
= 0.00428)
GOST 6651-94
–185 to 200
–301 to 392
± 0.72
± 1.296
Cu 100 (
= 0.00426)
GOST 6651-94
–50 to 200
–58 to 392
± 0.36
± 0.648
Cu 100 (
= 0.00428)
GOST 6651-94
–185 to 200
–301 to 392
± 0.36
± 0.648
Thermocouples
(2)
Type B
(3)
NIST Monograph 175, IEC 584
100 to 1820
212 to 3308
± 1.155
± 2.079
Type E
NIST Monograph 175, IEC 584
–200 to 1000
–328 to 1832
± 0.30
± 0.54
Type J
NIST Monograph 175, IEC 584
–180 to 760
–292 to 1400
± 0.525
± 0.945
Type K
(4)
NIST Monograph 175, IEC 584
–180 to 1372
–292 to 2501
± 0.75
± 1.35
Type N
NIST Monograph 175, IEC 584
–200 to 1300
–328 to 2372
± 0.75
± 1.35
Type R
NIST Monograph 175, IEC 584
0 to 1768
32 to 3214
± 1.125
± 2.025
Type S
NIST Monograph 175, IEC 584
0 to 1768
32 to 3214
± 1.05
± 1.89
Type T
NIST Monograph 175, IEC 584
–200 to 400
–328 to 752
± 0.525
± 0.945
DIN Type L
DIN 43710
–200 to 900
–328 to 1652
± 0.525
± 0.945
DIN Type U
DIN 43710
–200 to 600
–328 to 1112
± 0.525
± 0.945
Type W5Re/W26Re
ASTM E 988-96
0 to 2000
32 to 3632
± 1.05
± 1.89
GOST L
GOST R 8.585-2001
–200 to 800
–328 to 1472
± 0.525
± 0.945
Other input types
Millivolt Input
–10 to 100 mV
± 0.0225 mV
2-, 3-, 4-wire Ohm Input
0 to 2000 ohms
± 0.675 ohm
1.
The published digital accuracy applies over the entire sensor input range. Digital output can be accessed by HART Communications or
Wireless
HART.
2.
Total digital accuracy for thermocouple measurement: sum of digital ac0.8 °C (cold junction accuracy).
3.
Digital accuracy for NIST Type B T/C is ±4.5 °C (±8.1 °F) from 100 to 300 °C (212 to 572 °F).
4.
Digital accuracy for NIST Type K T/C is ±1.05 °C (±1.895 °F) from –180 to –90 °C (–292 to –130 °F).
Содержание Rosemount X-well 648
Страница 2: ......
Страница 6: ...vi Reference Manual 00809 0200 4648 Rev BB Title Page July 2016 Title Page ...
Страница 10: ...4 Reference Manual 00809 0200 4648 Rev BB Contents July 2016 Contents ...
Страница 12: ...2 Reference Manual 00809 0200 4648 Rev BB Introduction July 2016 Introduction ...
Страница 36: ...26 Reference Manual 00809 0200 4648 Rev BB Configuration July 2016 Configuration ...
Страница 52: ...42 Reference Manual 00809 0200 4648 Rev BB Commissioning July 2016 Commissioning ...
Страница 62: ...52 Reference Manual 00809 0200 4648 Rev BB Operation and Maintenance July 2016 Operation and Maintenance ...
Страница 66: ...56 Reference Manual 00809 0200 4648 Rev BB Troubleshooting July 2016 Troubleshooting ...
Страница 89: ......