36
Reference Manual
00809-0200-4648, Rev BB
July 2016
Option 1
1.
Connect sensor wiring shield to the transmitter housing (only if the housing is
grounded).
2.
Ensure the transmitter housing is electrically isolated from the sensor wiring.
A. Sensor wires
B. Transmitter
C. Shield ground point
Option 2
1.
Ground sensor wiring shield at the sensor.
2.
Ensure the sensor wiring and shield is electrically isolated from the transmitter
housing.
A. Sensor wires
B. Transmitter
C. Shield ground point
Note
Always use facility recommended wiring practices.
A
B
C
A
B
C