System Reference Manual
370XA Gas Chromatograph
7P00370-H01
APRIL 2014
4 - 8
4.2.8
Start a single stream analysis run
1.
Go to the
Main Menu
screen. This can be done in one of two
ways:
•
From the
Home
screen, press the
Enter
key.
•
From any other screen, press the
Exit
or the
Enter
key.
2.
From the
Main Menu
screen, use the left or right arrow keys to
move to the
GC Control
menu.
3.
Press the down arrow to highlight the
Single Stream
command.
4.
Press the
Enter
key. The
Start Single Stream Analysis
screen
displays.
5.
Press the down arrow key to highlight the stream that you want
to analyze.
The
Purge stream for 60 seconds
feature allows sample gas
to flow through the sample loop for 60 seconds prior to
beginning the first analysis. The feature is checked by default.
The
Continuous operation
feature allows for the repeated
analysis of the selected stream. The feature is checked by
default.
6.
To check or clear the
Purge stream for 60 seconds
check box
or the
Continuous operation
check box, do the following:
(a.)
Press the down arrow key to move from the
Stream
list
box to the
Purge stream for 60 seconds
check box.
(b.)
To check or clear the
Purge stream for 60 seconds
check box, press the
Select/Enter
key.
(c.)
Press the down arrow key to move from the
Purge
stream for 60 seconds
check box to the
Continuous
operation
check box.
(d.) To
check
or clear the
Continuous operation
check box,
press the
Select/Enter
key.
Содержание Rosemount 370XA
Страница 1: ...Revision A 7P00370 H01 April 2014 370XA Gas Chromatograph System Reference Manual ...
Страница 24: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 2 9 2 7 Interior Maintainable Module ...
Страница 32: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 5 3 3 2 2 Dimensions ...
Страница 38: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 11 3 3 3 2 Dimensions ...
Страница 60: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 33 3 9 1 Terminal wiring diagram ...
Страница 137: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 5 8 ...
Страница 138: ...6 1 Section 6 Overhauling the maintainable module ...
Страница 177: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 40 Column 2 ...
Страница 178: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 41 Column3 ...
Страница 179: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 42 Column 4 ...
Страница 180: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 43 Sample Pre Heat Coil Sph ...
Страница 181: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 44 Carrier Pre Heat Coil Cph ...
Страница 182: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 45 Sample Loop SL ...
Страница 183: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 46 Restrictor Column R ...
Страница 193: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 6 7 2 1 The GC won t power up ...
Страница 194: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 7 7 2 2 Un normalized total error ...
Страница 195: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 8 ...
Страница 196: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 9 7 2 3 Carrier pressure ...
Страница 197: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 10 7 2 4 Heater won t reach set point ...
Страница 198: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 11 ...
Страница 199: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 12 ...
Страница 200: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 13 ...
Страница 201: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 14 7 2 5 Module validation fails ...
Страница 202: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 15 ...
Страница 203: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 16 7 2 6 Auto valve timing fails ...
Страница 204: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 17 ...
Страница 205: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 18 ...
Страница 206: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 19 ...
Страница 207: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 20 This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank ...