370XA Gas Chromatograph
System Reference Manual
APRIL 2014
7P00370-H01
5 - 3
If the dome is too tight, insert two screwdrivers or similar tools
into the grooves (1) at the top rim of the dome to provide
additional leverage when twisting the dome.
CAUTION
The grooves are designed to aid you in loosening the dome. Do not attempt to use
them to tighten the dome.
7.
Remove the insulation cap.
WARNING
The oven will be approximately 80 °C (176 °F) and hot to the touch.
8.
Loosen the four 4 mm (5/32-inch) hex screws holding the
module to the base.
You may hear the release of some trapped carrier and sample
gases. If the release is continuous, confirm that the sample,
calibration, carrier, and actuation gases are isolated.
Содержание Rosemount 370XA
Страница 1: ...Revision A 7P00370 H01 April 2014 370XA Gas Chromatograph System Reference Manual ...
Страница 24: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 2 9 2 7 Interior Maintainable Module ...
Страница 32: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 5 3 3 2 2 Dimensions ...
Страница 38: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 11 3 3 3 2 Dimensions ...
Страница 60: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 33 3 9 1 Terminal wiring diagram ...
Страница 137: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 5 8 ...
Страница 138: ...6 1 Section 6 Overhauling the maintainable module ...
Страница 177: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 40 Column 2 ...
Страница 178: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 41 Column3 ...
Страница 179: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 42 Column 4 ...
Страница 180: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 43 Sample Pre Heat Coil Sph ...
Страница 181: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 44 Carrier Pre Heat Coil Cph ...
Страница 182: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 45 Sample Loop SL ...
Страница 183: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 46 Restrictor Column R ...
Страница 193: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 6 7 2 1 The GC won t power up ...
Страница 194: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 7 7 2 2 Un normalized total error ...
Страница 195: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 8 ...
Страница 196: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 9 7 2 3 Carrier pressure ...
Страница 197: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 10 7 2 4 Heater won t reach set point ...
Страница 198: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 11 ...
Страница 199: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 12 ...
Страница 200: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 13 ...
Страница 201: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 14 7 2 5 Module validation fails ...
Страница 202: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 15 ...
Страница 203: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 16 7 2 6 Auto valve timing fails ...
Страница 204: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 17 ...
Страница 205: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 18 ...
Страница 206: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 19 ...
Страница 207: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 20 This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank ...