System Reference Manual
370XA Gas Chromatograph
7P00370-H01
APRIL 2014
7 - 4
Auto Valve Timing Failure
The GC has taken too many analysis runs during an Auto
Valve Timing (AVT) routine without finding the ideal valve
timing. Confirm that the Calibration Gas isolation valves
are on and there is greater than 10 PSI (0.7 Bar) of
calibration gas. Refer to
“Auto valve timing fails” on
page 7-16
for help with troubleshooting this issue.
Analytical Module Communication
Failure
The Intelligent Module board on the analytical module is
not communicating with the CPU board.
Analytical Module Not Initialized
The analytical module failed to initialize when power was
applied.
Module Validation Failure
The analysis of the calibration gas failed during the
Module Validation process. Refer to
“Module validation
fails” on page 7-14
for help with troubleshooting this
issue.
IMB Incompatible
The firmware revision of the IMB is newer than the
firmware revision of the CPU board. Upgrade the CPU
firmware.
IMB CDT Component is missing in GC
A component that is in the component data table stored in
a recently installed module is not configured in the GC's
existing component data table. The component data table
of the GC must match (with the exception of neo-
Pentane) the component data table of a replacement
module.
Energy Value Invalid
If enabled, the GC analyzes the calibration gas as an
unknown stream and computes its energy value. The GC
then compares this value to the Cal Gas Cert CV and
determines if the calibration gas'energy value is within the
CV Check Allowed Deviation. If it isn't, the GC triggers the
Energy Value Invalid alarm.
Stream 1 Ideal RF Deviation
Streams 1 to 3 should only be set to Analysis or Not Used.
This alarm may occur If the streams are incorrectly
configured for calibration. To resolve this error, start
MON2020 and select
Streams
on the
Application
menu.
Set the stream’s
Usage
to
Analysis
or
Unused
.
Stream 2 Ideal RF Deviation
Stream 3 Ideal RF Deviation
Stream 4 Ideal RF Deviation
The response factor for one or more components is
outside of the limits of the Ideal Response Factor check
setting. Refer to the Failed Calibration Trouble-shooting
guide.
Alarm
Description and Suggested Actions
Содержание Rosemount 370XA
Страница 1: ...Revision A 7P00370 H01 April 2014 370XA Gas Chromatograph System Reference Manual ...
Страница 24: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 2 9 2 7 Interior Maintainable Module ...
Страница 32: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 5 3 3 2 2 Dimensions ...
Страница 38: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 11 3 3 3 2 Dimensions ...
Страница 60: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 33 3 9 1 Terminal wiring diagram ...
Страница 137: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 5 8 ...
Страница 138: ...6 1 Section 6 Overhauling the maintainable module ...
Страница 177: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 40 Column 2 ...
Страница 178: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 41 Column3 ...
Страница 179: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 42 Column 4 ...
Страница 180: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 43 Sample Pre Heat Coil Sph ...
Страница 181: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 44 Carrier Pre Heat Coil Cph ...
Страница 182: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 45 Sample Loop SL ...
Страница 183: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 46 Restrictor Column R ...
Страница 193: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 6 7 2 1 The GC won t power up ...
Страница 194: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 7 7 2 2 Un normalized total error ...
Страница 195: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 8 ...
Страница 196: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 9 7 2 3 Carrier pressure ...
Страница 197: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 10 7 2 4 Heater won t reach set point ...
Страница 198: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 11 ...
Страница 199: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 12 ...
Страница 200: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 13 ...
Страница 201: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 14 7 2 5 Module validation fails ...
Страница 202: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 15 ...
Страница 203: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 16 7 2 6 Auto valve timing fails ...
Страница 204: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 17 ...
Страница 205: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 18 ...
Страница 206: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 19 ...
Страница 207: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 20 This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank ...