System Reference Manual
370XA Gas Chromatograph
7P00370-H01
APRIL 2014
7 - 2
Warm Start Failed
The GC has failed to return to analysis mode after a
power failure. If the GC is in Auto Analysis mode and the
power is lost, when power is reapplied the unit will heat up
and analyze calibration gas until all of the component
peaks are identified. When all of the peaks are identified
and the analysis is good, the GC will go back into Auto
Analysis mode. However, if the peaks are not identified
within the time period indicated in the Max Warm Start
Delay field on the System window, the GC will raise the
Warm Start Fail alarm and go into Idle Mode and will not
analyze stream gas.
Heater 1 Out Of Range
The Analytical Oven Heater failed to reach the set-point
within 15 minutes or reads higher than the setpoint. Refer
to the Heater Failure troubleshooting guide.
Heater 2 Out Of Range
For the default configuration of the 370XA, this is not
used. The
Switch
setting on the Heaters screen for
Heater 2 must be set to
Fixed ON
and the
Fixed PWM
Output
set to
0
.
Detector 1 Scaling Factor Failure
The scaling factor generated for the high and low gain
preamp channel was outside the acceptable range of
values. This is typically caused by a fault on the analyzer
board, which should be replaced.
Carrier Pressure Low
The carrier gas pressure, as measured on the carrier gas
pressure sensor on the analytical module, is too low.
Check the regulated pressure at the bottle is at 90 PSI
(6.2 BarG), and confirm the isolation valves between the
bottle and the 370XA are all open. Refer to
“Carrier
pressure” on page 7-9
for help with troubleshooting this
issue.
Carrier Pressure Out Of Range
The carrier gas pressure, as measured on the carrier gas
pressure sensor on the analytical module, is outside of
the control limits. Check the regulated pressure at the
bottle is at 90 PSI (6.2 BarG). If the regulated pressure is
OK, check the connections on the Carrier Pressure
Control Valve and the pressure sensor. Refer to
“Carrier
pressure” on page 7-9
for help with troubleshooting this
issue.
Alarm
Description and Suggested Actions
Содержание Rosemount 370XA
Страница 1: ...Revision A 7P00370 H01 April 2014 370XA Gas Chromatograph System Reference Manual ...
Страница 24: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 2 9 2 7 Interior Maintainable Module ...
Страница 32: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 5 3 3 2 2 Dimensions ...
Страница 38: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 11 3 3 3 2 Dimensions ...
Страница 60: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 33 3 9 1 Terminal wiring diagram ...
Страница 137: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 5 8 ...
Страница 138: ...6 1 Section 6 Overhauling the maintainable module ...
Страница 177: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 40 Column 2 ...
Страница 178: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 41 Column3 ...
Страница 179: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 42 Column 4 ...
Страница 180: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 43 Sample Pre Heat Coil Sph ...
Страница 181: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 44 Carrier Pre Heat Coil Cph ...
Страница 182: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 45 Sample Loop SL ...
Страница 183: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 46 Restrictor Column R ...
Страница 193: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 6 7 2 1 The GC won t power up ...
Страница 194: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 7 7 2 2 Un normalized total error ...
Страница 195: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 8 ...
Страница 196: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 9 7 2 3 Carrier pressure ...
Страница 197: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 10 7 2 4 Heater won t reach set point ...
Страница 198: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 11 ...
Страница 199: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 12 ...
Страница 200: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 13 ...
Страница 201: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 14 7 2 5 Module validation fails ...
Страница 202: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 15 ...
Страница 203: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 16 7 2 6 Auto valve timing fails ...
Страница 204: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 17 ...
Страница 205: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 18 ...
Страница 206: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 19 ...
Страница 207: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 20 This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank ...