370XA Gas Chromatograph
System Reference Manual
APRIL 2014
7P00370-H01
7 - 3
Analog Input 1 High Signal
The value of the variable assigned to the analog output is
outside the range of the analog output. If the value is
valid, re-range the Analog Output and the associated
device that is receiving the signal so that the value of the
variable is within the range.
Analog Input 1 Low Signal
Analog Output 1 High Signal
Analog Output 1 Low Signal
Analog Output 2 High Signal
Analog Output 2 Low Signal
Stream 1 Validation Failure
The analysis of the validation gas for the associated
stream is outside the allowable percent deviation, as
defined in the Validation Data table. Confirm that the
Validation Gas concentration is entered correctly in the
validation data table. If the concentrations are entered
correctly, check the validation run analysis results and
chromatogram for analysis errors.
Stream 2 Validation Failure
Stream 3 Validation Failure
Stream 4 Validation Failure
Stream 5 Validation Failure
Stream 1 RF Deviation
Streams 1 to 3 should only be set to Analysis or Not Used.
This alarm may occur If the streams are incorrectly
configured for calibration. To resolve this error, start
MON2020 and select
Streams
on the
Application
menu.
Set the stream’s
Usage
to
Analysis
or
Unused
.
Stream 2 RF Deviation
Stream 3 RF Deviation
Stream 4 RF Deviation
The response factor for one or more components have
changed during a calibration run by a percentage greater
than set in the Component Data Table. Check that the
calibration gas concentration has been entered into the
component data table correctly, and confirm that the
calibration gas bottle is not empty or isolated. If the
concentrations are correct, the isolation valves are open,
and there is sufficient calibration gas pressure (higher
than 20 PSI or 1.4 Bar), check the calibration analysis and
chromatograms for analytical errors.
Analyzer Board Comm Failure
The communication link between the CPU board and the
analyzer board has failed. Remove power and confirm the
analyzer board is seated correctly on the back plane. If
the analyzer board is seated correctly, replace the
analyzer board. If replacing the analyzer board does not
rectify the issue, replace the back plane board.
LOI Comm Failure
The communication link between the CPU Board and the
Local Operator Interface (LOI) has failed. Remove power
and confirm the connectors from the LOI are seated
correctly in the back plane. If the connectors are seated
correctly, replace the LOI. If replacing the LOI does not
rectify the issue, replace the back plane board.
Alarm
Description and Suggested Actions
Содержание Rosemount 370XA
Страница 1: ...Revision A 7P00370 H01 April 2014 370XA Gas Chromatograph System Reference Manual ...
Страница 24: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 2 9 2 7 Interior Maintainable Module ...
Страница 32: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 5 3 3 2 2 Dimensions ...
Страница 38: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 11 3 3 3 2 Dimensions ...
Страница 60: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 3 33 3 9 1 Terminal wiring diagram ...
Страница 137: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 5 8 ...
Страница 138: ...6 1 Section 6 Overhauling the maintainable module ...
Страница 177: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 40 Column 2 ...
Страница 178: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 41 Column3 ...
Страница 179: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 42 Column 4 ...
Страница 180: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 43 Sample Pre Heat Coil Sph ...
Страница 181: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 44 Carrier Pre Heat Coil Cph ...
Страница 182: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 6 45 Sample Loop SL ...
Страница 183: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 6 46 Restrictor Column R ...
Страница 193: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 6 7 2 1 The GC won t power up ...
Страница 194: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 7 7 2 2 Un normalized total error ...
Страница 195: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 8 ...
Страница 196: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 9 7 2 3 Carrier pressure ...
Страница 197: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 10 7 2 4 Heater won t reach set point ...
Страница 198: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 11 ...
Страница 199: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 12 ...
Страница 200: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 13 ...
Страница 201: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 14 7 2 5 Module validation fails ...
Страница 202: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 15 ...
Страница 203: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 16 7 2 6 Auto valve timing fails ...
Страница 204: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 17 ...
Страница 205: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 18 ...
Страница 206: ...370XA Gas Chromatograph System Reference Manual APRIL 2014 7P00370 H01 7 19 ...
Страница 207: ...System Reference Manual 370XA Gas Chromatograph 7P00370 H01 APRIL 2014 7 20 This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank ...