Table 4-2: Startup Screen Sequence
(continued)
Sequence
Description
Screen
7. Secondary variable
Measurement value of mapped
second variable.
8. Electronics temperature
Temperature value of device
electronics.
9. Supply voltage
Voltage reading of the power
module.
> 6 V
Good
5.2 V – 6.0 V
Low
< 5.2 V
Very low
10. Percent of range
Level value in percent of total
measurement range.
11. Active alert present
This screen will only appear if there
is at least one active alert present.
For detailed information of the
failure source and recommended
actions, go to the
Active Alerts
screen in AMS Wireless Configurator
or the handheld communicator.
Refer to
Some active alerts will be displayed
on the LCD display as part of the
diagnostic button screen sequence,
refer to
.
Reference Manual
Configuration
00809-0100-4308
February 2020
Reference Manual
83
Содержание Rosemount 3308A
Страница 8: ...Contents Reference Manual February 2020 00809 0100 4308 viii Rosemount 3308 Series ...
Страница 10: ...Introduction Reference Manual February 2020 00809 0100 4308 10 Rosemount 3308 Series ...
Страница 22: ...Transmitter overview Reference Manual February 2020 00809 0100 4308 22 Rosemount 3308 Series ...
Страница 74: ...Installation Reference Manual February 2020 00809 0100 4308 74 Rosemount 3308 Series ...
Страница 96: ...Configuration Reference Manual February 2020 00809 0100 4308 96 Rosemount 3308 Series ...
Страница 194: ...Product Certifications Reference Manual February 2020 00809 0100 4308 194 Rosemount 3308 Series ...
Страница 222: ...Configuration parameters Reference Manual February 2020 00809 0100 4308 222 Rosemount 3308 Series ...