Figure A-5: Accuracy over Measuring Range for Flexible Twin Lead Probe
0.4 in. (1 cm)
3.5 in. (9 cm)
0.4 in. (1 cm)
3.5 in. (9 cm)
7.9 in.
(20 cm)
7.9 in.
(20 cm)
5.9 in. (15 cm)
9.5 in. (24 cm)
5.9 in. (15 cm)
9.5 in. (24 cm)
3.9 in. (10 cm)
12.6 in. (32 cm)
3.9 in. (10 cm)
9.8 in. (25 cm)
±0.12 in.
(3 mm)
±1.18 in.
(30 mm)
±0.12 in.
(3 mm)
±1.18 in.
(30 mm)
±0.2 in.
(5 mm)
±1.18 in.
(30 mm)
±0.2 in.
(5 mm)
±1.18 in.
(30 mm)
E
E
E
E
C
D
C
D
A
B
F
A. High performance (profile code U)
B. Standard (profile code S)
C. Water (DC = 80)
D. Oil (DC = 2.2)
E. Accuracy
F. Blind Zone
Specifications and reference data