Table 6-6: Incorrect level readings
(continued)
Symptom
Possible cause and recommended actions
Level stuck in measuring range.
•
May be caused by a disturbing object in the tank. Read
the Echo Curve and adjust Thresholds, see
.
•
Check if the probe is bent and in contact with the tank
wall. This contact causes a false echo reading.
•
Heavy coating or contamination on the probe. Clean
the probe.
Level stuck in empty.
•
Thresholds may be too high, see
Level fluctuations (a couple of inches).
•
There is too much disturbing noise in the tank (from
foam, splashing etc.). Set the Performance Mode to
High to get a stable measurement signal, refer to
Resolve noise or weak surface echoes
•
There are rapid level changes in the tank. Select a faster
Update Rate, refer to
•
Thin oil layer on top of water that is sometimes
detected, sometimes not. Set the Peak Detection
Method to Threshold Intersection to improve the
stability of level measurements in such applications.
Refer to
.
Table 6-7: Incorrect or missing interface level reading
Symptom
Possible cause and recommended actions
Incorrect Interface Level reading.
•
The Upper Product Dielectric Constant is not correct,
see
Upper product dielectric constant
.
•
Air gap is too big when the Measurement Mode is set to
Interface Level with Submerged Probe. Reduce air gap
or switch the Measurement Mode to Product Level and
Interface Level.
Missing Interface Level reading (reported as NaN).
•
The Measurement Mode is set to Product Level. Set
Measurement Mode to Product Level and Interface
Level, refer to
.
•
The Interface Peak is difficult to detect, because the
bottom product has a low dielectric constant, or the
signal is attenuated in the upper product. Check
Thresholds. For more information, see
There are two products in the tank, but only the product
surface or interface is detected.
•
The upper product is too thin to be detected. No action
is needed. See
interface thickness for different probe types.
•
Check Thresholds, see
There are two products in the tank, but no readings are
reported.
•
Check Thresholds, see
Service and troubleshooting
Reference Manual
February 2020
00809-0100-4308
116
Rosemount 3308 Series
Содержание Rosemount 3308A
Страница 8: ...Contents Reference Manual February 2020 00809 0100 4308 viii Rosemount 3308 Series ...
Страница 10: ...Introduction Reference Manual February 2020 00809 0100 4308 10 Rosemount 3308 Series ...
Страница 22: ...Transmitter overview Reference Manual February 2020 00809 0100 4308 22 Rosemount 3308 Series ...
Страница 74: ...Installation Reference Manual February 2020 00809 0100 4308 74 Rosemount 3308 Series ...
Страница 96: ...Configuration Reference Manual February 2020 00809 0100 4308 96 Rosemount 3308 Series ...
Страница 194: ...Product Certifications Reference Manual February 2020 00809 0100 4308 194 Rosemount 3308 Series ...
Страница 222: ...Configuration parameters Reference Manual February 2020 00809 0100 4308 222 Rosemount 3308 Series ...