Register transducer block
The Register Transducer Block allows access to Database registers and Input registers of the Rosemount
™
2240S Multi-input Temperature Transmitter. This makes it possible to read a selected set of register
directly by accessing the memory location.
The Register Transducer Block is only available with advanced service.
Table C-5. Register Transducer Block Parameters
Since the Register Transducer Block allows access to most registers in the Rosemount 2240S, which
includes registers set by the Methods and Configuration screens in the Measurement Transducer Block
(see
“Measurement transducer block” on page 169
), it should be handled with care and ONLY to be
changed by trained and certified service personnel, or as guided by Emerson Automation Solutions/
Rosemount Tank Gauging support personnel.
Index
Number
Parameter
Description
1
ST_REV
The revision level of the static Bdata associated with the
function block. The revision value increments each time a
static parameter value in the block is changed.
2
TAG_DESC
The user description of the intended application of the
block.
3
STRATEGY
The strategy field can be used to identify grouping of
blocks. This data is not checked or processed by the block.
4
ALERT_KEY
The identification number of the plant unit. This
information may be used in the host for sorting alarms, etc.
5
MODE_BLK
The actual, target, permitted, and normal modes of the
block.
Target: The mode to “go to”
Actual: The mode the “block is currently in”
Permitted: Allowed modes that target may take on
Normal: Most common mode for target
6
BLOCK_ERR
This parameter reflects the error status associated with the
hardware or software components associated with a block.
It is a bit string, so that multiple errors may be shown.
7
UPDATE_EVT
This alert is generated by any change to the static data
8
BLOCK_ALM
The block alarm is used for all configuration, hardware,
connection failure or system problems in the block. The
cause of the alert is entered in the subcode field. The first
alert to become active will set the Active status in the Status
parameter. As soon as the Unreported status is cleared by
the alert reporting task, another block alert may be
reported without clearing the Active status, if the subcode
has changed.
9
TRANSDUCER_DIRECTORY
Directory that specifies the number and starting indices of
the transducers in the transducer block.
10
TRANSDUCER_TYPE
Identifies the transducer.
11
TRANSDUCER_TYPE_VER
12
XD_ERROR
A transducer block alarm sub code.
Содержание Rosemount 2240S
Страница 2: ......
Страница 10: ...viii Reference Manual 00809 0100 2240 Rev DA Title Page October 2017 Title Page ...
Страница 16: ...6 Reference Manual 00809 0100 2240 Rev DA Introduction October 2017 Introduction ...
Страница 28: ...18 Reference Manual 00809 0100 2240 Rev DA Overview October 2017 Overview ...
Страница 116: ...106 Reference Manual 00809 0100 2240 Rev DA Configuration Operation October 2017 Configuration Operation ...
Страница 150: ...140 Reference Manual 00809 0100 2240 Rev DA Service and Troubleshooting October 2017 Service and Troubleshooting ...
Страница 164: ...154 Product Certifications Product Certifications October 2017 Reference Manual 00809 0100 2240 Rev DA ...
Страница 192: ...182 Reference Manual 00809 0100 2240 Rev DA Index October 2017 Index ...
Страница 193: ......