158
October 2017
Reference Manual
00809-0100-2240, Rev DA
43
FD_FAIL_ACTIVE
This parameter reflects the error conditions that are being
detected as active as selected for this category. It is a bit string, so
that multiple conditions may be shown.
44
FD_OFFSPEC_ACTIVE
45
FD_MAINT_ACTIVE
46
FD_CHECK_ACTIVE
47
FD_FAIL_MAP
This parameter maps conditions to be detected as active for this
alarm category. Thus the same condition may be active in all,
some, or none of the 4 alarm categories.
48
FD_OFFSPEC_MAP
49
FD_MAINT_MAP
50
FD_CHECK_MAP
51
FD_FAIL_MASK
This parameter allows the user to suppress any single or multiple
conditions that are active, in this category, from being broadcast to
the host through the alarm parameter. A bit equal to ‘1’ will mask
i.e. inhibit the broadcast of a condition, and a bit equal to ‘0’ will
unmask i.e. allow broadcast of a condition.
52
FD_OFFSPEC_MASK
53
FD_MAINT_MASK
54
FD_CHECK_MASK
55
FD_FAIL_ALM
This parameter is used primarily to broadcast a change in the
associated active conditions, which are not masked, for this alarm
category to a Host System.
56
FD_OFFSPEC_ALM
57
FD_MAINT_ALM
58
FD_CHECK_ALM
59
FD_FAIL_PRI
This parameter allows the user to specify the priority of this alarm
category.
60
FD_OFFSPEC_PRI
61
FD_MAINT_PRI
62
FD_CHECK_PRI
63
FD_SIMULATE
This parameter allows the conditions to be manually supplied when
simulation is enabled. When simulation is disabled both the
diagnostic simulate value and the diagnostic value tracks the
actual conditions. The simulate jumper is required for simulation to
be enabled and while simulation is enabled the recommended
action will show that simulation is active.
64
FD_RECOMMEN_ACT
This parameter is a device enumerated summarization of the most
severe condition or conditions detected. The DD help should
describe by enumerated action, what should be done to alleviate
the condition or conditions. 0 is defined as Not Initialized, 1 is
defined as No Action Required, all others defined by manufacturer.
65
FD_EXTENDED_ACTIVE
An optional parameter or parameters to allow the user finer detail
on conditions causing an active condition in the FD_*_ACTIVE
parameters.
66
FD_EXTENDED_MAP
An optional parameter or parameters to allow the user finer control
on enabling conditions contributing to the conditions in
FD_*_ACTIVE parameters.
67
COMPATIBILITY_REV
This parameter is used when replacing field devices. The correct
value of this parameter is the DEV_REV value of the replaced
device.
68
HARDWARE_REVISION
Hardware revision.
69
SOFTWARE_REV
Software revision of source code with resource block.
Index
Number
Parameter
Description
Содержание Rosemount 2240S
Страница 2: ......
Страница 10: ...viii Reference Manual 00809 0100 2240 Rev DA Title Page October 2017 Title Page ...
Страница 16: ...6 Reference Manual 00809 0100 2240 Rev DA Introduction October 2017 Introduction ...
Страница 28: ...18 Reference Manual 00809 0100 2240 Rev DA Overview October 2017 Overview ...
Страница 116: ...106 Reference Manual 00809 0100 2240 Rev DA Configuration Operation October 2017 Configuration Operation ...
Страница 150: ...140 Reference Manual 00809 0100 2240 Rev DA Service and Troubleshooting October 2017 Service and Troubleshooting ...
Страница 164: ...154 Product Certifications Product Certifications October 2017 Reference Manual 00809 0100 2240 Rev DA ...
Страница 192: ...182 Reference Manual 00809 0100 2240 Rev DA Index October 2017 Index ...
Страница 193: ......