102
HART Communications
Section 10: haRT Communication
Instruction Manual
April 2017
LIQ-MAN-1066
1066DO Device variables
1066Oz Device variables
1066Cl Device variables
Device variable Name
assignable to Dynamic variables
variable Range
primary value Type:
Oxygen (17)
PV, SV, TV or QV
0 to 100 ppm
0 to 1000 ppb
0 to 100 mg/L
0 to 1000 µg/L
0 to 300 % Saturation
0 to 760 mmHg
0 to 30 inHg
0 to 1 bar
0 to 1000 mbar
0 to 100 kPa
0 to 1 atm
Other Device variables:
Temperature
PV, SV, TV or QV
-15 to 200 °C
5 to 360 °F
Temperature resistance
TV or QV
0 to 100000 Ω
Sensor input current
PV, SV, TV or QV
0 to 100000 nA
Device variable Name
assignable to Dynamic variables
variable Range
primary value Type:
Ozone (18)
PV, SV, TV or QV
0 to 20 ppm
0 to 1000 ppb
0 to 20 mg/L
0 to 1000
μ
g/L
Other Device variables:
Temperature
PV, SV, TV or QV
-15 to 200 °C
5 to 360 °F
Temperature resistance
TV or QV
0 to 100000
μ
Sensor input current
PV, SV, TV or QV
0 to 100000 nA
Device variable Name
assignable to Dynamic variables
variable Range
primary value Type:
Chlorine (19)
PV, SV, TV or QV
0 to 20 ppm
0 to 1000 ppb
0 to 20 mg/L
0 to 1000 µg/L
Other Device variables:
Temperature
PV, SV, TV or QV
-15 to 200 °C
5 to 360 °F
Temperature resistance
TV or QV
0 to 100000 Ω
Sensor input current
PV, SV, TV or QV
0 to 100000 nA
pH (free chlorine pH compensation) SV, TV or QV
0 to 14 pH
Содержание Rosemount 1066
Страница 1: ...Instruction Manual LIQ MAN 1066 Rev J April 2017 Rosemount 1066 Smart Enabled 2 Wire Transmitter ...
Страница 2: ......
Страница 10: ...VIII Table of Contents Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 22: ...12 Specifications Section 2 Description and Specifications Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 26: ...16 Installation Section 3 Installation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 36: ...26 Wiring Section 4 Wiring Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 46: ...36 Display and Operation Section 6 Display and Operation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 76: ...66 Programming Measurements Section 8 programming Measurements Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 130: ...120 HART Communications Section 10 haRT Communication Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 132: ...EC Declaration of Conformity Instruction Manual April 2017 LIQ MAN 1066 122 Return of Material ...