Wiring
25
Instruction Manual
Section 4: Wiring
LIQ-MAN-1066
April 2017
OUTPUT 2
TB5
TB3
TB2
TB4
TB1
+24V
GND
GND
+24V
T
H
U
M
ANODE
CATHODE
RTN
SNS
RTD IN
+V
-V
REF
SHLD
SENSOR WIRING
TB7
TB6
_
+
GND
SOL
SHLD
pH
(OUTPUT1)
LOOP PWR
4-
20
m
A
/ -2
4V
D
C
RE
TU
RN
4-2
0m
A
/ +
24
V
D
C
INSTALL PLUGS IN ALL OTHER
OPENINGS AS NEEDED
INNER ENCLOSURE
HINGE SIDE OF FRONT PANEL
1
TB7
/OUTPUT 2 REQUIRES EXTERNAL DC POWER.
2
TB6/
THUM TERMINAL IS USED ONLY FOR
WIRELESS THUM ADAPTOR INSTALLATIONS
.
1
2
1066 HART CIRCUIT BOARD
(pH/CL/DO/OZ)
ASSY 24406-xx
HINGED PANEL
DWG NO.
40106613
FIGuRE 4-10. haRT loop power Wiring
Содержание Rosemount 1066
Страница 1: ...Instruction Manual LIQ MAN 1066 Rev J April 2017 Rosemount 1066 Smart Enabled 2 Wire Transmitter ...
Страница 2: ......
Страница 10: ...VIII Table of Contents Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 22: ...12 Specifications Section 2 Description and Specifications Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 26: ...16 Installation Section 3 Installation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 36: ...26 Wiring Section 4 Wiring Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 46: ...36 Display and Operation Section 6 Display and Operation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 76: ...66 Programming Measurements Section 8 programming Measurements Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 130: ...120 HART Communications Section 10 haRT Communication Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 132: ...EC Declaration of Conformity Instruction Manual April 2017 LIQ MAN 1066 122 Return of Material ...