Section 8: programming Measurements
Instruction Manual
April 2017
LIQ-MAN-1066
52
Programming Measurements
Slope
The display screen for Entering the conductivity/temp Slope is shown.
The default value is displayed in bold type. Refer to the toroidal
conductivity Programming flow diagram to complete this function.
Ref Temp
The display screen for manually Entering the Reference temperature is
shown. The default value is displayed in bold type. Refer to the toroidal
conductivity Programming flow diagram to complete this function.
Filter
The display screen for entering the input filter value in seconds is shown.
The default value is displayed in bold type. Refer to the toroidal
conductivity Programming flow diagram to complete this function.
using the highest range (90 mS to 1500 mS) in very low conductivity
processes below 100 µS (conductance) might generate a high noise
value relative to the actual process value. In these cases, it is
recommended to increase the input filter setting above the default
value of 2 sec. to suppress the effect of noise.
Custom Setup
The display screens for creating custom curves for converting
conductivity to concentration is shown. Refer to the toroidal
conductivity Programming flow diagram to complete this function.
When the custom curve data entry is complete, press ENTER. The
display will confirm the determination of a custom curve fit to the
entered data by displaying this screen:
If the custom curve fit is not completed or is unsuccessful, the display will
read as follows and the screen will return to the beginning custom curve
screen.
8.5.7
8.5.8
8.5.9
8.5.10
1.234 µS/cm 25.0 ºC
S
N
Slope
2.00%/ºC
1.234 µS/cm 25.0 ºC
S
N
Ref Temp
(25.0ºC normal)
+25.0ºC
1.234 µS/cm 25.0 ºC
S
N
Input filter
02 sec
1.234 µS/cm 25.0 ºC
S
N
Custom Curve
Configure
Enter Data Points
Calculate Curve
1.234 µS/cm 25.0 ºC
S
N
Calculate Curve
Custom curve
fit completed.
In Process Cal
recommended.
1.234 µS/cm 25.0 ºC
S
N
Calculate Curve
Failure
Содержание Rosemount 1066
Страница 1: ...Instruction Manual LIQ MAN 1066 Rev J April 2017 Rosemount 1066 Smart Enabled 2 Wire Transmitter ...
Страница 2: ......
Страница 10: ...VIII Table of Contents Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 22: ...12 Specifications Section 2 Description and Specifications Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 26: ...16 Installation Section 3 Installation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 36: ...26 Wiring Section 4 Wiring Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 46: ...36 Display and Operation Section 6 Display and Operation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 76: ...66 Programming Measurements Section 8 programming Measurements Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 130: ...120 HART Communications Section 10 haRT Communication Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 132: ...EC Declaration of Conformity Instruction Manual April 2017 LIQ MAN 1066 122 Return of Material ...