Instruction Manual
Section 9: Calibration
LIQ-MAN-1066
April 2017
Calibration
79
Calibration – Monochloromine
A monochloramine sensor generates a current directly proportional to the concentration of
monochloramine in the sample. Calibrating the sensor requires exposing it to a solution
containing no monochloramine (zero standard) and to a solution containing a known amount of
monochloramine (full-scale standard). The zero calibration is necessary because monochloramine
sensors, even when no monochloramine is in the sample, generate a small current called the
residual or zero current. The transmitter compensates for the residual current by subtracting it
from the measured current before converting the result to a monochloramine value. New sensors
require zeroing before being placed in service, and sensors should be zeroed whenever the
electrolyte solution is replaced. The best zero standard is deionized water. The purpose of the In
Process calibration is to establish the slope of the calibration curve. Because stable
monochloramine standards do not exist, the sensor must be calibrated against a test run on a grab
sample of the process liquid. Several manufacturers offer portable test kits for this purpose.
To calibrate the chlorine sensor, access the Calibration screen by pressing ENTER from the main
screen, and press ENTER.
The following calibration routines are covered:
1. Zero Cal Zeroing the sensor in solution with zero total chlorine
2. Grab Cal Standardizing to a sample of known monochloramine concentration
To calibrate monochloramine:
1. Press the MENU button
2. Select Calibrate. Press ENTER.
3. Select Monochloramine. Press ENTER.
The adjacent screen will appear. To calibrate Monochloramine or
Temperature, scroll to the desired item and press ENTER.
The following sub-sections provide you with the initial display screen that appears for each
calibration routine. Use the flow diagram for Chlorine calibration at the end of Sec. 9 and the live
screen prompts to complete calibration.
The adjacent screen appears after selecting Monochloramine calibration:
9.6.3
1.234 µS/cm 25.0 ºC
S
N
Calibration
Zero Cal
In Process Cal
1.234 µS/cm 25.0 ºC
S
N
Calibrate?
Monochloramine
Temperature
Содержание Rosemount 1066
Страница 1: ...Instruction Manual LIQ MAN 1066 Rev J April 2017 Rosemount 1066 Smart Enabled 2 Wire Transmitter ...
Страница 2: ......
Страница 10: ...VIII Table of Contents Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 22: ...12 Specifications Section 2 Description and Specifications Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 26: ...16 Installation Section 3 Installation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 36: ...26 Wiring Section 4 Wiring Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 46: ...36 Display and Operation Section 6 Display and Operation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 76: ...66 Programming Measurements Section 8 programming Measurements Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 130: ...120 HART Communications Section 10 haRT Communication Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 132: ...EC Declaration of Conformity Instruction Manual April 2017 LIQ MAN 1066 122 Return of Material ...