96
HART Communications
Section 10: haRT Communication
Instruction Manual
April 2017
LIQ-MAN-1066
Measurements available via haRT
A number of live measurements are made available by HART in addition to the main
measurements such as pH or Conductivity. All of these measurements are called Device Variables,
which can be mapped to the Dynamic Variables PV, SV, TV, and QV for regular reading by the
typical HART host.
Each 1066 transmitter type, 1066P, 1066C, etc. will have its own set of Device Variables, based on
the secondary measurements used in making the main measurement.
shows the
Device Variable for the each transmitter type, and the Dynamic Variables, which they can be
mapped to.
Diagnostics available via haRT
Status Information – Device Status Bits
Bit 0 primary variable out of limits:
This bit is set when PV is out of its limits.
Bit 1
Non-primary variable out of limits:
This bit is set when any active device variable other than the Primary variable is out of its
limits.
Bit 2
loop Current Saturated:
This bit is set when Analog Output 1 is not fixed, and it is less than 3.8 mA or greater than
22.0 mA.
Bit 3
loop Current Fixed:
This bit is set when Analog Output 1 is being simulated, calibrated, or when a device
failure is detected and the Analog Output 1 is configured to output a fixed value.
Bit 4
More Status available:
The “more status available” bit will be set when the device status condition occurs (i.e. bit
goes from 0 to 1) on at least one of the Additional Transmitter Status bits are set.
Bit 5 Cold Start:
This bit is set when a Master Reset is performed either by Command 42, or a power cycle.
2 bits are maintained internally, for primary and secondary masters.
Bit 6
Configuration Changed:
This bit is set when a configuration or calibration parameter is changed either through a
write command or a local interface command. 2 bits are maintained internally, for
primary and secondary masters.
Bit 7
Field Device Malfunction:
This bit is set when a fault condition is detected in the device electronics or sensor.
10.4
10.4.1
10.3
Содержание Rosemount 1066
Страница 1: ...Instruction Manual LIQ MAN 1066 Rev J April 2017 Rosemount 1066 Smart Enabled 2 Wire Transmitter ...
Страница 2: ......
Страница 10: ...VIII Table of Contents Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 22: ...12 Specifications Section 2 Description and Specifications Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 26: ...16 Installation Section 3 Installation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 36: ...26 Wiring Section 4 Wiring Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 46: ...36 Display and Operation Section 6 Display and Operation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 76: ...66 Programming Measurements Section 8 programming Measurements Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 130: ...120 HART Communications Section 10 haRT Communication Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 132: ...EC Declaration of Conformity Instruction Manual April 2017 LIQ MAN 1066 122 Return of Material ...