8
Specifications
Section 2: Description and Specifications
Instruction Manual
April 2017
LIQ-MAN-1066
Recommended Sensors for Conductivity
All Rosemount 400 series conductivity sensors (Pt 1000 RTD) and 410VP 4-electrode sensor.
Toroidal Conductivity (Codes
–
T)
Measures conductivity in the range of 1 µS/cm to 2,000,000 µS/cm (2 S/cm). Measurement choices
are conductivity, resistivity, total dissolved solids, salinity, and % concentration. The % concentration
selection includes the choice of five common solutions (0-12% NaOH, 0-15% HCl, 0-20% NaCl, and 0-
25% or 96-99.7% H
2
SO
4
). The conductivity concentration algorithms for these solutions are fully
temperature compensated. For other solutions, a simple-to-use menu allows the customer to enter
his own data. The transmitter accepts as many as five data points and fits either a linear (two points)
or a quadratic function (three to five points) to the data. Reference temperature and linear tempera-
ture slope may also be adjusted for optimum results. Two temperature compensation options are
available: manual slope (X% / °C) and neutral salt (dilute sodium chloride). Temperature compensa-
tion can be disabled, allowing the transmitter to display raw conductivity. For more information con-
cerning use of the toroidal conductivity sensors, refer to the product data sheets.
Performance Specifications
Temperature specifications:
Repeatability:
±0.25% ±5 µS/cm after
2.4.2
2.5
2.5.1
Temperature range
-25 to 210 °C (-13 to 410 °F)
Temperature Accuracy,
Pt-100, -25 to 50 °C
± 0.5 °C
Temperature Accuracy,
Pt-100, 50 to 210 °C
± 1 °C
226: ±1% of reading ±5
µ
S/cm in recommended range
225 & 228: ±1% of reading ±15
µ
S/cm in recommended range
222, 242: ±4% of reading ±5mS/cm in recommended range
225, 226 & 228: ±5% of reading outside high recommended range
loop performance
(Following Calibration)
TaBlE 2-2. performance Specifications: Recommended Range – Toroidal Conductivity
High performance 225 Toroidal &
226 Conductivity sensors
Model 1
µ
S/cm 10
µ
S/cm 100
µ
S/cm 1000
µ
S/cm 10mS/cm 100mS/cm 1000mS/cm 2000mS/cm
50
µ
S/cm to 500mS/cm
50
µ
S/cm to 1500mS/cm
500mS/cm to 2000mS/cm
500
µ
S/cm to 2000mS/cm
100
µ
S/cm to 2000mS/cm
1500mS/cm to 2000mS/cm
226
242
222
(1in & 2in)
225 & 228
Содержание Rosemount 1066
Страница 1: ...Instruction Manual LIQ MAN 1066 Rev J April 2017 Rosemount 1066 Smart Enabled 2 Wire Transmitter ...
Страница 2: ......
Страница 10: ...VIII Table of Contents Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 22: ...12 Specifications Section 2 Description and Specifications Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 26: ...16 Installation Section 3 Installation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 36: ...26 Wiring Section 4 Wiring Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 46: ...36 Display and Operation Section 6 Display and Operation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 76: ...66 Programming Measurements Section 8 programming Measurements Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 130: ...120 HART Communications Section 10 haRT Communication Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 132: ...EC Declaration of Conformity Instruction Manual April 2017 LIQ MAN 1066 122 Return of Material ...