Intrinsically Safe Installation
29
Instruction Manual
Section 5: Intrinsically Safe Installation
LIQ-MAN-1066
April 2017
FIGuRE 5-3. CSa Installation
OUTPUT2
TB2
TB1
LOOP PWR
DSHLD
DRV_A
RTN
DRV_B
RSHLD
RCV_A
SHLD
RTDIN
TB7
TB6
GND
+24V
THUM
RCV_B
SENS
E
+24V
OUTPUT2
TB2
TB1
LOOP PWR
DSHLD
DRV_A
RTN
DRV_B
RSHLD
RCV_A
SHLD
RTDIN
TB7
TB6
GND
GND
+24V
THUM
RCV_B
SENS
E
OUTPUT 2
(OUTPUT1)
LOOP PWR
TB5
TB3
TB2
TB4
TB1
GND
GND
THUM
ANODE
CATHODE
RTN
SNS
RTD IN
REF
SHLD
SENSOR WIRING
TB7
TB6
GND
SOL
SHLD
WARNING- SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY OR
SUITABILITY FOR DIVISION 2.
WARNING- TO PREVENT IGNITION OF FLAMMABLE OR COMBUSTIBLE ATMOSPHERES,
DISCONNECT POWER BEFORE SERVICING.
ROSEMOUNT MODEL 375 OR 475
FIELD COMMUNICATOR REMOTE
TRANSMITTER INTERFACE FOR USE
IN CLASS I AREA
(SEE NOTE 3 AND TABLE III)
LOAD
IS CLASS I, GRPS A-D
CLASS II, GRPS E-G
CLASS III
NI CLASS I, DIV 2
GRPS A-D
CLASS II, DIV 2
GRPS E-G
ANALOG OUTPUT 2 ONLY
AVAILABLE ON 1066...HT...
APPROVED MODEL 222, 225, 226 OR 228 TOROIDAL CONDUCTIVITY SENSOR
OR
MODELS 140, 141, 142, 150, 400, 400VP, 401, 402, 402VP, 403, 403VP, 404 &
410VP CONTACTING CONDUCTIVITY SENSOR
OR
ANY SIMPLE APPARATUS WITH LESS THAN 200 FEET OF ATTACHED CABLE
UNSPECIFIED POWER SUPPLY
30 VDC MAX FOR INTRINSIC
SAFETY (24 VDC TYPICAL)
17.5 VDC MAX. FOR FISCO OPTION
SAFETY BARRIER
(SEE NOTES 2 & 8 FOR FISCO,
SEE NOTES 2, 8, 9 & 10
FOR ALL OTHER OPTIONS.)
LOAD
UNSPECIFIED POWER SUPPLY
30 VDC MAX FOR INTRINSIC
SAFETY (24 VDC TYPICAL)
17.5 VDC MAX. FOR FISCO OPTION
SAFETY BARRIER
(SEE NOTES 2 & 8 FOR FISCO,
SEE NOTES 2, 8, 9 & 10
FOR ALL OTHER OPTIONS.)
PH SENSOR
CSA APPROVED DEVICE OR
SIMPLE APPARATUS
AMPEROMETRIC SENSOR
CSA APPROVED DEVICE OR
SIMPLE APPARATUS
1066-CL/DO/OZ...ONLY
EITHER OR BOTH
MAY BE INSTALLED
ANALOG OUTPUT 2 ONLY AVAILABLE ON 1066...HT...
ANALOG OUTPUT 2 ONLY AVAILABLE ON 1066....HT...
OR
APPROVED MODEL 222, 225, 226 OR 228 TOROIDAL CONDUCTIVITY SENSOR
OR
MODELS 140, 141, 142, 150, 400, 400VP, 401, 402, 402VP, 403, 403VP, 404 &
410VP CONTACTING CONDUCTIVITY SENSOR
OR
ANY SIMPLE APPARATUS WITH LESS THAN 200 FEET OF ATTACHED CABLE
1066C... ENTITY OUTPUT PARAMETERS: Voc = 4.75V, Isc = 676.93mA, Pmax = 258mW
THE 1066C IS CERTIFIED AS A SYSTEM FOR USE WITH MODELS 140, 141, 142, 150, 400, 400VP, 401, 402, 402VP, 403, 403VP, 404 & 410VP
CONTACTING
CONDUCTIVITY SENSORS, OR ANY SIMPLE APPARATUS WITH < 200 FEET OF CABLE.
THE 1066T... IS CERTIFIED AS A SYSTEM FOR USE WITH THE MODELS 222, 225, 226 & 228 TOROIDAL CONDUCTIVITY SENSORS, OR ANY SIMPLE
APPARATUS
WITH < 200 FEET OF CABLE.
17
18
17
18
15
16
17
15
16
17
15
16
17
15
16
17
17
18
REV
E
DWG NO
D
Содержание Rosemount 1066
Страница 1: ...Instruction Manual LIQ MAN 1066 Rev J April 2017 Rosemount 1066 Smart Enabled 2 Wire Transmitter ...
Страница 2: ......
Страница 10: ...VIII Table of Contents Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 22: ...12 Specifications Section 2 Description and Specifications Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 26: ...16 Installation Section 3 Installation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 36: ...26 Wiring Section 4 Wiring Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 46: ...36 Display and Operation Section 6 Display and Operation Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 76: ...66 Programming Measurements Section 8 programming Measurements Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 130: ...120 HART Communications Section 10 haRT Communication Instruction Manual April 2017 LIQ MAN 1066 ...
Страница 132: ...EC Declaration of Conformity Instruction Manual April 2017 LIQ MAN 1066 122 Return of Material ...