Configuration and Use Manual
55
Configuring Volumetric Flow Measurement
V
o
lumetr
ic
Flo
w
Measure
ment
Histor
y
, Events, a
n
d Alar
ms
I/
O
Core Interfa
c
e
Figure 6-4
Advanced panel
Table 6-4
Advanced parameters
Parameter
Description
Comments
Base Pressure
Enter the pressure at which gas base volume will be calculated
(i.e., the pressure to which gas base volume will be corrected).
Base
Temperature
Enter the temperature at which gas base volume will be
calculated (i.e., the temperature to which gas base volume will
be corrected).
Pressure Tap
Type
• Gauge: Pressure is read from a gauge. The pressure value
will be corrected for atmospheric pressure.
• Absolute: The pressure input provides absolute pressure
values. No correction will be applied.
Atmospheric
Pressure
• Calculate: Atmospheric pressure will be calculated according
to the site elevation. Enter the site elevation in the textbox
provided.
• Enter: The configured atmospheric pressure will be used.
Enter the atmospheric pressure value in the textbox provided.
Atmospheric pressure data is
required only if Pressure Tap Type
is set to Gauge.
Содержание Micro Motion 3711
Страница 32: ...24 Micro Motion Model 3711 Gas Flow Computers ...
Страница 44: ...36 Micro Motion Model 3711 Gas Flow Computers ...
Страница 56: ...48 Micro Motion Model 3711 Gas Flow Computers ...
Страница 68: ...60 Micro Motion Model 3711 Gas Flow Computers ...
Страница 92: ...84 Micro Motion Model 3711 Gas Flow Computers ...
Страница 108: ...100 Micro Motion Model 3711 Gas Flow Computers ...
Страница 140: ...132 Micro Motion Model 3711 Gas Flow Computers ...
Страница 196: ...188 Micro Motion Model 3711 Gas Flow Computers ...
Страница 210: ...202 Micro Motion Model 3711 Gas Flow Computers ...
Страница 233: ......