e
rface
A005
Input Overrange
Check the test points. See Section 15.16.
Check the sensor pins. See Section 15.18.
Verify process.
Verify calibration factors in Model 3711 configuration. See Section 15.13.
Re-zero the Model 3711. See Section 5.6.
A008
Density Overrange
Check the test points. See Section 15.16.
Check the sensor pins. See Section 15.18.
Verify process. Check for air in the flow tubes, tubes not filled, foreign material
in tubes, or coating in tubes.
Verify Model 3711 characterization. See Section 15.13.
Perform density calibration. See Section B.4.
A010
Calibration Failure
If alarm appears during a zero procedure, ensure that there is no flow through
the sensor, then retry.
Cycle power to the meter, then retry.
A016
Line Temp Out-of-range
Check the test points. See Section 15.16.
Check the sensor pins. See Section 15.18.
Check wiring to sensor. See Section 15.15.2.
Contact Micro Motion. See Section 1.8.
A017
Meter Temp
Out-of-Range
Check the test points. See Section 15.16.
Check the sensor pins. See Section 15.18.
Contact Micro Motion. See Section 1.8.
A020
Cal Factor Unentered
Check the characterization. Specifically, verify the FCF value. See
Section 15.13.
A021
Incorrect Sensor Type
Check the characterization. Specifically, verify the K1 value. See Section 15.13.
A022
Configuration Corrupt
Cycle power to the meter.
The Model 3711 might need service. Contact Micro Motion. See Section 1.8.
A023
Totals Corrupt
Cycle power to the meter.
The Model 3711 might need service. Contact Micro Motion. See Section 1.8.
A024
CP Program Corrupt
Cycle power to the meter.
The Model 3711 might need service. Contact Micro Motion. See Section 1.8.
A025
Boot Sector Fault
Cycle power to the meter.
The Model 3711 might need service. Contact Micro Motion. See Section 1.8.
A102
Drive Overrange
Excessive drive gain. See Section 15.16.2.
Check the sensor pins. See Section 15.18.
A104
Calibration in Progress
Allow the meter to complete calibration.
A105
Slug Flow
A107
Power Reset
No action required.
NA
Sensor Not Vibrating
Contact Micro Motion. See Section 1.8.
NA
Low Amplitude
Contact Micro Motion. See Section 1.8.
NA
Clip Flag
Contact Micro Motion. See Section 1.8.
Table 15-1
Status alarms and remedies
continued
Alarm
code
Message
Possible remedy
Содержание Micro Motion 3711
Страница 32: ...24 Micro Motion Model 3711 Gas Flow Computers ...
Страница 44: ...36 Micro Motion Model 3711 Gas Flow Computers ...
Страница 56: ...48 Micro Motion Model 3711 Gas Flow Computers ...
Страница 68: ...60 Micro Motion Model 3711 Gas Flow Computers ...
Страница 92: ...84 Micro Motion Model 3711 Gas Flow Computers ...
Страница 108: ...100 Micro Motion Model 3711 Gas Flow Computers ...
Страница 140: ...132 Micro Motion Model 3711 Gas Flow Computers ...
Страница 196: ...188 Micro Motion Model 3711 Gas Flow Computers ...
Страница 210: ...202 Micro Motion Model 3711 Gas Flow Computers ...
Страница 233: ......