200
Micro Motion
®
Model 3711 Gas Flow Computers
Calibration
During calibration, the sensor must be completely filled with the calibration fluid, and flow through
the sensor must be at the lowest rate allowed by your application. This is usually accomplished by
closing the shutoff valve downstream from the sensor, then filling the sensor with the appropriate
fluid.
B.4.2
Calibration procedure
To calibrate for density:
1. Click
Meter > Calibration
and open the Density panel. The dialog box shown in Figure B-4 is
displayed.
2. Follow the procedure illustrated in Figure B-5.
Note: ROCLINK 800 automatically switches to Auto Scan mode so that the messages will be updated
automatically. This affects only the calibration procedure. When you close the window, Auto Scan will
be returned to its previous setting.
3. If you need to cancel the density calibration process, click the appropriate
Cancel Calibration
button. The Model 3711 will use the pre-existing calibration values.
Figure B-4
Density Calibration panel
Содержание Micro Motion 3711
Страница 32: ...24 Micro Motion Model 3711 Gas Flow Computers ...
Страница 44: ...36 Micro Motion Model 3711 Gas Flow Computers ...
Страница 56: ...48 Micro Motion Model 3711 Gas Flow Computers ...
Страница 68: ...60 Micro Motion Model 3711 Gas Flow Computers ...
Страница 92: ...84 Micro Motion Model 3711 Gas Flow Computers ...
Страница 108: ...100 Micro Motion Model 3711 Gas Flow Computers ...
Страница 140: ...132 Micro Motion Model 3711 Gas Flow Computers ...
Страница 196: ...188 Micro Motion Model 3711 Gas Flow Computers ...
Страница 210: ...202 Micro Motion Model 3711 Gas Flow Computers ...
Страница 233: ......