Table 2-5: 1815B Zero crossing preamplifier performance
(continued)
Parameter type
Description
Outputs
Powered pulse output:
•
Type: Square wave
•
Frequency range: ≤ 5 kHz
•
Amplitude: 0-12 V
Construction
Delrin housing
Solid epoxy encapsulation
Table 2-6: 1817X Bi-directional preamplifier performance
Parameter type
Description
Inputs
Supply voltage: 11.5-36 VDC
Sensor type: Reluctance
Signal: Sine wave
Preamplifier sensitivity: 40 mVpp
Temperature range: -40 ºC to 85 ºC (-40 ºF to 185 ºF)
Outputs
Powered pulse output:
•
Type: Square wave
•
Frequency range: ≤ 5 kHz
•
Amplitude: 0-Vs
Construction
Delrin housing
Solid epoxy encapsulation
Wiring and cable connections
The pickoff input signal connections are made at terminal strip connection TB2 for
Channel A, and TB3 for Channel B. Power supply and output signal connections are made
at terminal strip TB1. Refer to
.
Important
•
For single channel wiring use multi-conductor control and instrumentation cable
(18AWG), or equivalent.
•
Earth ground shield at one end only.
•
Insulate the shield at the other end.
•
For dual channel wiring use multi-conductor control and instrumentation cable
(18AWG), or equivalent.
•
Connect both shields to earth ground at one end of the cables and insulate the shields
at the other end.
•
The LME housing should be at earth ground.
Operating conditions and specifications
User manual
March 2019
P/N 3-9008-515
48
Daniel Series 1500 Liquid Turbine Meter, NPS 3-24
Содержание Daniel Series
Страница 1: ...User manual P N 3 9008 515 Rev AF March 2019 Daniel Series 1500 Liquid Turbine Meter NPS 3 through 24 ...
Страница 2: ...2 ...
Страница 11: ...Part I Plan User manual Plan P N 3 9008 515 March 2019 User manual 11 ...
Страница 12: ...Plan User manual March 2019 P N 3 9008 515 12 Daniel Series 1500 Liquid Turbine Meter NPS 3 24 ...
Страница 69: ...Part II Install User manual Install P N 3 9008 515 March 2019 User manual 69 ...
Страница 70: ...Install User manual March 2019 P N 3 9008 515 70 Daniel Series 1500 Liquid Turbine Meter NPS 3 24 ...
Страница 78: ...Testing User manual March 2019 P N 3 9008 515 78 Daniel Series 1500 Liquid Turbine Meter NPS 3 24 ...
Страница 79: ...Part III Operate User manual Operate P N 3 9008 515 March 2019 User manual 79 ...
Страница 80: ...Operate User manual March 2019 P N 3 9008 515 80 Daniel Series 1500 Liquid Turbine Meter NPS 3 24 ...
Страница 83: ...Part IV Maintain User manual Maintain P N 3 9008 515 March 2019 User manual 83 ...
Страница 84: ...Maintain User manual March 2019 P N 3 9008 515 84 Daniel Series 1500 Liquid Turbine Meter NPS 3 24 ...
Страница 105: ...User manual P N 3 9008 515 March 2019 User manual 105 ...