Table 2-3: 2818 Dual channel preamplifier performance
(continued)
Parameter type
Description
Variable voltage output:
•
Type: Square wave
•
Frequency range: 0 to 5 kHz
•
Amplitude: 0 to Supply voltage
•
1000 Ohm internal pull-ups, 70 mA, max.
Construction
Delrin housing
Solid epoxy encapsulation
Table 2-4: 2415 Intrinsically safe preamplifier performance
Parameter type
Description
Inputs
Supply voltage: 13-25 Vdc @ ≤12 mA
Sensor type: Reluctance
Signal: Sine wave
Preamplifier sensitivity:
•
60 mVpp
•
10Hz - 1KHz (60mVpp)
•
10Hz - 3KHz (115mVpp)
•
10Hz - 5KHz (200mVpp)
Temperature range: -40 ºC to 85 ºC (-40 ºF to 185 ºF)
Outputs
Powered pulse output:
•
Type: Square wave
•
Frequency range: ~5 Hz to 5 kHz
•
Amplitude: 0 to 5 V
Construction
Delrin housing
Solid epoxy encapsulation
Table 2-5: 1815B Zero crossing preamplifier performance
Parameter type
Description
Inputs
Supply voltage: 14-30 VDC
Sensor type: Reluctance
Signal: Sine wave
Preamplifier sensitivity: 40 mVpp
Temperature range: -40 ºC to 85 ºC (-40 ºF to 185 ºF)
User manual
Operating conditions and specifications
P/N 3-9008-515
March 2019
User manual
47
Содержание Daniel Series
Страница 1: ...User manual P N 3 9008 515 Rev AF March 2019 Daniel Series 1500 Liquid Turbine Meter NPS 3 through 24 ...
Страница 2: ...2 ...
Страница 11: ...Part I Plan User manual Plan P N 3 9008 515 March 2019 User manual 11 ...
Страница 12: ...Plan User manual March 2019 P N 3 9008 515 12 Daniel Series 1500 Liquid Turbine Meter NPS 3 24 ...
Страница 69: ...Part II Install User manual Install P N 3 9008 515 March 2019 User manual 69 ...
Страница 70: ...Install User manual March 2019 P N 3 9008 515 70 Daniel Series 1500 Liquid Turbine Meter NPS 3 24 ...
Страница 78: ...Testing User manual March 2019 P N 3 9008 515 78 Daniel Series 1500 Liquid Turbine Meter NPS 3 24 ...
Страница 79: ...Part III Operate User manual Operate P N 3 9008 515 March 2019 User manual 79 ...
Страница 80: ...Operate User manual March 2019 P N 3 9008 515 80 Daniel Series 1500 Liquid Turbine Meter NPS 3 24 ...
Страница 83: ...Part IV Maintain User manual Maintain P N 3 9008 515 March 2019 User manual 83 ...
Страница 84: ...Maintain User manual March 2019 P N 3 9008 515 84 Daniel Series 1500 Liquid Turbine Meter NPS 3 24 ...
Страница 105: ...User manual P N 3 9008 515 March 2019 User manual 105 ...