28
Transducer cable removal and installation
Section 3: Meter maintenance
Reference, Installation and Operation Manual
December 2017
3-9000-764 Rev F
The meter body ports are identified by stamped or cast lettering adjacent to the transducer port
(i.e. Model 3814 - A1, A2, B1, B2, C1, C2, D1, and D2.
1.
Remove power to the meter.
2.
If installed, remove the security wire seals from the transducer cable connector
attached to the transducer cable nut and the transducer retainer attached to the
locking ring.
Figure 3-16 Transducer security seal removal
3.
Unscrew the cable nut from the transducer housing and then, pull the cable from the
transducer housing (
).
CUTTING HAZARD
Sharp edges may be present on the meter.
Wear appropriate personal protective equipment when working on the meter. Failure to do so may cause
serious injury.
A.
B.
A. Transducer cable connector to transducer cable nut security wire
B. Transducer retainer to locking ring security wire