Transducer cable removal and installation
Section 3: Meter maintenance
Reference, Installation and Operation Manual
December 2017
3-9000-764 Rev F
32.
If required, attach the security wire seals on the Base Enclosure.
Figure 3-20 Base Enclosure security seals
33.
If required by the site operations manager, have an electrician fully test the
connections. After the Acceptance Test is witnessed and approved, seal the conduit.
34.
Power down the system and apply the sealing compound to the conduit and allow to
set in accordance with manufacturer specifications.
35.
This completes the transducer cable removal and installation procedure.