32
Transducer cable removal and installation
Section 3: Meter maintenance
Reference, Installation and Operation Manual
December 2017
3-9000-764 Rev F
23.
Inspect the Transmitter Electronics Enclosure gasket for wear and replace it if necessary.
24.
If replacing the gasket, lubricate it with Dow Corning 111
1
(P/N 2-9-9960-135).
25.
Attach one desiccant pack to the underside of the Base Enclosure cover.
26.
Place the Transmitter Electronics Enclosure onto the Base Enclosure. Rotate the
Transmitter Electronics Enclosure until the mounting holes are correctly aligned with
the holes in the Base Enclosure.
27.
Install the two screws with a 6 mm Allen wrench to secure the Transmitter Electronics
Enclosure to the Base Enclosure.
28.
Reattach the external ground wire to the ground lug, reconnect the conduit and
prepare to seal the meter.