16
Transducer removal and installation
Section 3: Meter maintenance
Reference, Installation and Operation Manual
December 2017
3-9000-764 Rev F
Supplies needed:
•
Dow Corning 111
1
Silicone compound
•
Dow Corning 200
1
Silicon oil
•
Loctite® Nickel anti-seize compound
1.
Before removing and installing transducer(s), connect to the meter using Daniel
MeterLink and collect and save a Maintenance Log and configuration files.
2.
If installed, remove the security wire seals from the transducer cable connector
attached to the transducer cable nut and the transducer retainer attached to the
locking ring.
Figure 3-7 Transducer security seal removal
3.
Do not remove the transducer housing locking ring to ensure that the transducer
housing does not unscrew from the meter body during this process. Place a wrench on
the 1inch hex of the transducer retainer and unscrew it from the transducer housing.
4.
Use a pair of needle nose pliers to grab the indexing key of the transducer and pull the
transducer from the housing.
5.
For LT-05 transducers only, place a small amount of the Dow Corning 111
1
Silicone
grease (P/N 1-360-01-651) on the face of the transducer to be installed. Using your
finger spread the silicone grease over the face of the transducer to create a thin, even
layer of silicone grease on the surface of the transducer face. There should be no gaps or
mounds of the silicone grease on the face of the transducer. Wipe off any excess silicone
grease from the sides of the transducer.
1.
Dow Corning 111 and
Dow Corning 200 are trademarks of Dow Corning Corporation, U.S.A.
A.
A. Transducer cable connector to transducer cable nut security wire
B. Transducer retainer to locking ring security wire
B.
DO NOT ATTEMPT TO REMOVE TRANSDUCER HOUSING LOCKING RING
Removing the locking ring may allow the transducer housing to unscrew from the
meter body while the meter is under pressure.
Release of fluid under pressure may cause injury to personnel resulting in injury to personnel
or damage equipment.