8
Collect As Found logs and reports
Section 2: Setup meter communications
3410 Series Gas Ultrasonic Meters Operations Manual
January 2018
3-9000-777 Rev E
2.1.3
Collect As Found logs and reports
Use the Daniel MeterLink
Logs/Reports
menu and collect and save the logs and reports for a
historical record of the meter a t several velocities within the operating range of the meter to
establish a baseline to be used for the trending of the meter diagnostics and save the following
logs and reports for the “As Found” settings (factory default).
•
Maintenance log
•
System log
•
Save your meter’s configuration file
–
Use the
Tools|Edit/Compare Configuration
menu to display the dialog. Enable the
View All or Metrology
radio button and click
Read
.
The
All View
displays the meter’s extended configuration parameters. Grayed out
values are read-only, but help describe the configuration. The
Metrology View
displays the path dimensions, pipe diameter, transducer delay parameters, zero flow
calibration coefficients, and flow calibration coefficients.
Click the question mark icon for the data point to display additional information.
–
Click
Save
to save the configuration file. By default, the file is saved to the Data Folder
setup in
File|Program Settings
.