© ElektroPhysik
QuintSonic 7
177 von 184
Error message
Problem
Remedy
No printer / PC found !
Possible causes:
-
poor IR connection.
Check the direct visual
contact between the IR
windows. For data
transfer, an unobstructed
view between the IR
devices must be ensured.
Remove any error
sources.
-
empty storage batteries in
printer
-
No application software
launched on the PC
-
A wrong terminal device
has been chosen.
Ensure an unobstructed
visual contact between gauge
and printer.
Check storage battery state in
the printer, recharge or
replace batteries if necessary.
Launch the application
software on your PC.
Choose the appropriate
terminal device.
Data transfer error !
Failure during transfer of data
to printer or PC. Possible
causes:
poor IR connection. Check
the direct visual contact
between the IR windows. For
data transfer, an unobstructed
view between the IR devices
must be ensured. Remove
any error sources.
- empty storage batteries in
printer.
Ensure an unobstructed
visual contact between gauge
and printer.
Check storage battery state in
the printer, recharge or
replace batteries if necessary.
The following errors may be remedied by a total reset (see section 16.1)
-
keyboard failure
-
no further measurements possible
-
illogical readings
If no key action is possible or if the gauge cannot be switched off via the ON/OFF button, briefly
remove batteries and reinsert to remedy.