Group
Parameters & description
DC output
State after OT alarm
Determines the condition of the DC output after an overtemperature (OT) alarm, once the device has
cooled down:
•
Off
= DC output will remain off
•
Auto
= Default, the device will automatically restore the situation before the OT alarm, which usually
means the DC output to be on
Master-slave
Mode
Selecting
Master
or
Slave
enables the master-slave mode (MS) and defines the position for the unit in
the MS system. For details see section
“3.11.1. Parallel operation in master-slave (MS)“”.
Termination resistor
Activates or deactivates the so-called bus termination of the digital master-slave bus via a switchable
resistor. Termination should be activation if required, usually when problems with the master-slave bus
operation occur.
Bias resistors
Additionally to the regular termination resistor (TERM) this activates two bias resistor, if required, to help
stabilize the bus further. Tap on the information symbol for a graphical depiction.
Backlight off after 60s
If activated, it will switch off the display’s backlight after 60 seconds of inactivity. This settings is pri-
marily intended for slave units where the display isn’t supposed to be permanently on. It’s identical to
the setting in menu
HMI setup
.
Initialize system
Tapping this touch area will repeat the initialization of the master-slave system in case the detection
of all slave units by the master was unsuccessful, so the system would have less total power than ex-
pected, or has to be repeated manually in case the master unit couldn’t detect a missing slave or one
slave has failed.
USB logging
Log file separator format
Defines the format of CSV files generated from logging files (also see
and
3.4.8). This setting
also affects other features where a CSV file can be loaded or saved.
•
US
= Comma as column separator (US standard for CSV files)
•
Default
= Semicolon as column separator (german/european standard for CSV files)
Logging with units (V,A,W)
CSV files generated from USB logging by default add physical units to values. This can be deactivated
here.
USB logging
Activates/deactivates logging to USB stick. For more information refer to
“3.4.8. Recording to USB stick
Logging interval
Defines the time between two records in the log file. Selection:
500 ms
,
1 s
,
2 s
,
5 s
Start/stop
Defines how the USB logging is started and stopped.
•
Manual
= Logging only starts and stops upon user interaction on the HMI, by accessing touch button
in the quick menu.
•
At DC on/off
= Logging starts and stops with every change of state on the DC output, no matter if
caused by the user, software or a device alarm. Attention: Every next start will create a new log file.