8. HOW TO USE THE INSTRUMENT ELBO CONTROLLI srl
Preset E236+
41
8.8 HOW TO RESTORE STANDARD SETTINGS
E236+ stores all settings (from display and axes origin settings to system settings: optical calibration and
measurement compensation) in an EEPROM memory.
All these data are saved in two different memory areas and completed by the CRC control to check their
integrity. Indeed, it may occur that stored data are altered or are not complete due to power supply faults
during information storage or due to external electromagnetic events (lighting, etc.)
Any t is switched on, it controls data integrity, and, in case of alteration, it initialises the
concerned memory area once more.
Therefore, after the starting screen, which displays the software release, it may happen that the wording is
displayed
Specify the automatic deletion of the axes origin and display data area and the system settings area.
In the latter case, carry out the vision system calibration procedure (para. 8.7.1), under pain of measurement
inaccuracy.
In case the instrument behaves in an irregular way, you can force memory re-initialization by restoring the
standard settings:
Turn off the device by disconnecting the power supply and then connect it again and keep the central area
of the screen pressed until the start-up message disappears. Doing this, the content of the memory will be
fully initialised.
A first message will indicate the recovery process is underway
after the standard settings are restored, you will be asked if you want to cancel the camera calibration values
and then to perform the camera calibration procedure (para.8.7.1).