4-1. SelfCalibration functions
Set the details for "SelfCalibration".
Function
Settings
Description
Execute
-
SelfCalibration can be executed manually
regardless of the schedule.
Note
•
After selecting "Execute", warming up (after
the power to the monitor is switched on,
leaving the monitor on for a certain period of
time until the display becomes stable) may be
executed before the built-in calibration sensor
comes out.
Se
ttings
Mode Settings Standard Mode
On
Off
Switch between enabling/disabling
SelfCalibration in Standard Mode.
Advanced
Mode
CAL1
CAL2
CAL3
On
Off
Switch between enabling/disabling
SelfCalibration in Advanced Mode.
Schedule
Start time
Power Save
Immediately
Application
Off
Select the timing to execute SelfCalibration upon
reaching the time set in the schedule.
•
"Power Save"
To be executed under any of the following
conditions.
- When the monitor is in "Power Save" mode
or power is off at the set time.
- The monitor shifts to the power saving
mode or the power is turned off when the
set time in the schedule is exceeded.
•
"Immediately"
SelfCalibration is executed immediately at
the set time.
•
"Application"
Execute SelfCalibration according to the
timing set using ColorNavigator Network. For
details on ColorNavigator Network, refer to
our web site (
•
"Off"
SelfCalibration is not executed.
Frequency
Daily
Weekly
Monthly
Quarterly
Biannually
Annually
Usage Time
Select the SelfCalibration execution cycle.
36