4 Installing
the E-connect S
4. Installing the E-connect S
4.1 Installation of the
contra angle
Make sure 4 pins on contra angle
alignment the slots of handpiece, plug
them together until it “click” securely
into place.
The contra angle can be 340 degrees
rotated without take off, make it easy to
watch the LCD in treatment by rotating
the contra-angle.
WARNING
●
Make sure the assembly is
connected properly, otherwise might
cause unexpected motor reverse, even
hurt the patients
●
After connecting the contra angle and
handle, pull it gently to make sure the
connection is good.
The improve insulation of the contra
angle during combine apex, we
recommend using an insulating sleeve.
You can also use disposable sleeve
(sold separately) instead of insulating
sleeve
NOTE
Without the insulating sleeve, when
performing the apex measurement with
handpiece, wear appropriate insulated
gloves, and make sure the contra angle
does not touch the lips. It is advisable
to use a rubber dam when performing
such treatments.
4.2 Install the file
Turn the file back and forth until it is
lined up with interior latch groove and
slips into place, lock the file into the
contra angle.
Hold down the push button on the
contra angle and can release the file.
Push
Button
Page 8 / 47