7 Operation
●
Electromagnetic noise in surroundings environment may interfere with the
device operation, do not rely on device’s automatic control completely, always pay
attention to the feedback from display.
NOTE
If there is any abnormal functioning, stop using the device and report to company.
The file separates more easily at high speeds, please follow the manufacturer’s
recommendations of the speed and check the settings of the Endo motor before
use.
Do not use the files are except nickel-titanium or stainless steel.
Gloves and a rubber dam are compulsory during treatment.
Do not forget to remove the file from the Contra-angle after its use.
3
1
05
2
A
P
When using motor combine apex
function, the measure wire must be
connecting with motor by USB
socket, and white slot connects with
lip by lip hook, keep the black slot
idle.
The reference point bar will show on
the screen
(more information about
reference point bar, please see
chapter 5.2 Screen display)
.
3
1
-2
2
AP
NOTE
We strongly recommend check the
function every time before use.
Touch the lip hook with the file in the
contra angle and check that all the
bars on the meter on the screen light
up, and the motor should be reversed
continuously.