7 Operation
NOTE
The file clip must hold the file correctly.
Push the button of file clip in the
direction shown by the arrow, clip the
holder onto the metal upper part of the
file and then release the button.
The file clip must be nearly vertical with
file handle, otherwise will damage the
tip of the file holder.
mm
0.5
03
02
01
00
-1
04
NOTE
We recommend using Apex Tester to
check the accuracy of apex locator
every time before use.
Insert the Apex Tester into handpiece
USB socket during M0 mode, check the
highlight number must between 01-03
(0.3mm-0.8mm on the top)
OVER
0 3
0 2
0 1
0 0
- 1
0 4
NOTE
We recommend check the connection of
apex locator every time before use.
Touch the lip hook with metal part of file
clip, check all the bars on the meter on
the screen light up, and “over” flash on
the top.
Unsuitable situation of root canals for Electric Measurement
Cannot obtain precise measurements if the root canal conditions as below
Root canal with a large apical
foramen
The root canal cannot be accurately
measured because of the lesion or
incomplete development of the apical
foramen. The results may show that the
length measured is shorter than the
actual one.
Page 28 / 47