8 Cleaning
, Disinfection and Sterilization
STEP NO. INSTRUCTIONS
1
Initial
treatment at
point of use
Immediately after using, wipe gross contaminations
from the components (contra angle, file clip, lip
hook and insulating sleeve), and put them in
container for transportation.
Prepare the components directly after treatment.
●
Do not submerge the components or wipe them
with any of the following functional water (acidic
electrolyzed water, strong alkaline solution, or
ozone water), medical agents (glutaral, etc.), or
any other special types of water or commercial
cleaning liquids. Such liquids may result in
metal corrosion and adhesion of the residual
medical agents to the components.
2
Preparation
before
cleaning
Remove and disconnect the components (Contra
Angle, Lip Hook, File clip, Insulating Sleeve) before
cleaning. Refer to
"Chapter 4-Installing the
E-connect S" of this manual for disassembly
instructions.
WARNING
●
Do not fail to take out the file before cleaning the
contra angle.
●
Observe suitable personal protective measures.
The following Step 3 to Step 5 are operated in a washer-disinfector:
WARNING
●
Use only approved washer-disinfectors according to EN ISO 15883, maintain
and calibrate it regularly.
●
Follow instructions and observe concentrations given by the manufacturer (see
general recommendations).
●
Sufficient rinsing step should be available in purified water (max 10 germs/ml
and max 0.25 endotoxin units/ml)
●
Avoid any contact between the contra-angle and any instrument, kit, support or
container.
●
Make sure the components are dry before moving to the #6 step.
3
Cleaning:
Automated
- Carefully put the components (Contra Angle, Lip
Hook, File Clip, Insulating Sleeve) into the
washer-disinfector and set the parameters as
follows:
Page 33 / 47