8 Cleaning
, Disinfection and Sterilization
- Pre-cleaning: water temperature <30
℃
, 2 min;
- Cleaning: water temperature 45
℃
, 5 min; use
an enzyme detergent solution (mild and
aldehyde free solution) which is suitable to be
used with washer-disinfector, and use in
accordance with the IFU of the detergent
solution manufacturer;
- Rinsing: water temperature 45
℃
, 1 min (Rinsing
twice).
4
Disinfection:
Thermal
Thermal disinfection at least 5 min at 90
℃
/194
°
F,
make sure A0 value
≥
3000.
5
Drying
Heat, 20min, 90
℃
/194
°
F
6
Maintenance
and Inspection
-
Inspect components and sort out those with
defects. Dirty components must be cleaned and
disinfected again.
-
Only the contra angle needs to be lubricated.
Black oil
WARNING
●
Before autoclaving, the contra angle must be
lubricated.
●
Attaching the spray nozzle to oil can and contra
angle, press the oil can button more than 3
seconds, till all the black oil flow out from the
head of the contra angle.
7
Packaging
Pack each component in a separate
steam-sterilization pouch.
WARNING
●
Check the validity period of pouch given by the
manufacturer to determine the shelf life.
●
Use pouches which resist to a temperature up
Page 34 / 47