![Eighteeth E-connect S Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/eighteeth/e-connect-s/e-connect-s_user-manual_2369377016.webp)
6 Setting
NOTE
When choice REV mode, a slow beep alarm
sound appears after starting the motor, used for
indicating counter clockwise rotation happening.
Repeatedly
press
S
to check all the next level parameters of this operation
mode are expected, press
<
or
>
to change if not.
NOTE
The parameter will differ in difference mode according to certain logic
(See
chapter 6.5 Parameter logic)
.
Speed
300
rpm
The speed setting can be adjusted from 120 rpm
to 1000 rpm.
NOTE
The speed of REC and ATC operation mode is
difference according to certain logic
(See chapter
.
Torque L imit
3.0 N cm
·
The torque setting can be adjusted from 0.5 N·cm
to 4.0 N·cm, and R.L (torque reverse less) is also
available.
NOTE
The torque of REC and ATC operation mode is
difference according to certain logic
(See chapter
.
WARNING
When choice R.L (torque reverse less), a slow
beep alarm sound appears after starting the motor.
Be careful to use this function, very professional
skill is needed, otherwise has risk of file broken.
Aut o St art
ON
E-connect S
integrated apex locator, if the lip hook
is connecting with patient’s lip, when the endo file
entering root canal, the motor will start
automatically.
Press
<
or
>
to shut off this function if not
expected, press
●
to start and stop the motor.
WARNING
The motor will start automatically if handpiece
Page 16 / 47
Содержание E-connect S
Страница 1: ...0197 E connect S USER MANUAL...