5 Use
Interface
5.3
Terms and definition
Fwd
Forward ( Clockwise rotation )
Rev
Reverse ( Counter clockwise rotation )
Be applied to special file, inject calcium hydroxide and other
solutions
REC
Reciprocation
Be applied to reciprocating file, path file and rotary file
protection by setting some special angle
ATC
Adaptive torque control
Up to setting torque, the motor will move with reciprocating
mode; when torque reduce to normal value, the motor will
clockwise rotate
EAL
Electronic apex locator
In the mode, the device will work like a stand-alone apex
locator
AP
Apex
Major apical foramen or Anatomic apical foramen
R.L
Torque reverse less
The motor will not reverse rotation no matter how large the
torque load is
Reference point
During combined length determination, normally apical reverse
must active before reaching major apical foramen, setting
apical reverse position by change the flash bar
FWD Angle
Forward angle (Clockwise rotation angle), activating in REC
and ATC operation mode
REV Angle
Reverse angle (Counter Clockwise rotation angle), activating
in REC and ATC operation mode
Memory Mode
Such as M0-M10
Operation Mode Such as FWD, REV, REC and ATC