and then connect the electrical supply again and repeat the entire process.
Continue repeating until no further cooling water is purged from the system.
8.
The cooling water drain procedure is now complete. Disconnect the air supply,
electrical supply and the drain hoses. Dispose of the drained cooling water
appropriately.
Note:
An alternative method to drain the cooling water from the system is to:
- obey the procedure described above until step 6
- at Step 7 use the PDT to manually open all the system water valves (through the
Test mode menu), as described in
Connect the cooling water supply
on page 45.
9.4 Transportation
WARNING: OIL LEAKAGE
Do not drain the oil from the pump(s) if dangerous substances have been pumped or
not. Blanking plates must be installed to seal all vacuum inlet and outlet ports (to
prevent possible oil leakage). Make sure that the system is correctly labelled, if in
doubt contact the supplier.
WARNING: STORAGE SAFETY
Risk of injury or damage to equipment. Follow the pictorial symbols on outside of
packages for transport and storage limitation.
Follow the procedure given in
on page 89 and then read form HS1 and fill out
form HS2, which can be found at the back of printed manuals or can be downloaded
from
http:// edwardsvacuum.com/HSForms/
10/2020 - ©Edwards Limited
Page 88
M56635880_D
M56635880_D - Decommissioning
Содержание iXM Series
Страница 1: ...Dry Pump System iXM INSTRUCTION MANUAL M56635880_D Original instructions edwardsvacuum com...
Страница 101: ...This page has been intentionally left blank...
Страница 102: ...This page has been intentionally left blank...
Страница 103: ...This page has been intentionally left blank...
Страница 104: ...edwardsvacuum com...