4.6.2 Connect the pump inlet
CAUTION: O-RING SAFETY
Risk of damage to the equipment. For iXM200 and iXM3000 variants, do not use a
trapped O
‑
ring or centering ring for installation. Use 8 half claw clamps to attach inlet
piping and tighten it to a torque of 5.0 Nm.
1.
Remove the temporary cover or blanking plate from the inlet of the system. Do not
drop screws, tools etc. into the pump inlet.
A.
Keep the nuts, bolts, washers and blanking plate for future use.
B.
Keep the temporary cover for future use on non
‑
contaminated pumps only.
2.
Connect the inlet flange (
Figure: Front view of the system
) to the vacuum system.
A.
For iXM200 and iXM3000 variants, use the O
‑
ring (supplied) and applicable
nuts, bolts and washers (not supplied) to connect the inlet flange. The inlet
O
‑
ring is supplied with the pump and is installed underneath the inlet flange
cover. Refer to
Figure: Connect the pump inlet (iXM200 and iXM3000)
B.
For iXM600 and iXM1200 variants, use the trapped O
‑
ring or centre
ring(supplied) and applicable nuts, bolts and washers (not supplied) to
connect the inlet flange. Refer to
Figure: Connect the pump inlet (iXM600
3.
Use half claw clamps when you connect an ISO style foreline flange to the dry
pumping system inlet. Refer to
Figure: Connect the pump inlet
Figure 9 Connect the pump inlet (iXM200 and iXM3000)
10/2020 - ©Edwards Limited
Page 33
M56635880_D
M56635880_D - Installation
Содержание iXM Series
Страница 1: ...Dry Pump System iXM INSTRUCTION MANUAL M56635880_D Original instructions edwardsvacuum com...
Страница 101: ...This page has been intentionally left blank...
Страница 102: ...This page has been intentionally left blank...
Страница 103: ...This page has been intentionally left blank...
Страница 104: ...edwardsvacuum com...