Fault
Warning 190.01 - Valve Missing
Cause
The DP stator cooling solenoid valve is disconnected or missing (open circuit).
Remedy
Make sure that the DP stator cooling solenoid valve is installed and is correctly
connected.
Fault
Warning 190.01 - Valve Fault
Cause
The DP stator cooling solenoid valve is faulty (short circuit).
Remedy
Contact us.
Fault
Warning 191.01 - Valve Missing
Cause
The DP motor cooling solenoid valve is disconnected or missing (open circuit).
Remedy
Make sure that the DP motor cooling solenoid valve is installed and is correctly
connected.
Fault
Warning 191.01 - Valve Fault
Cause
The DP motor cooling solenoid valve is faulty (short circuit).
Remedy
Contact us.
Fault
Warning 192.01 - Valve Missing
Cause
The DP end cover cooling solenoid valve is disconnected or missing (open circuit)
Remedy
Make sure that the DP stator cooling solenoid valve is installed and is correctly
connected.
Fault
Warning 192.01 - Valve Fault
Cause
The DP end cover cooling solenoid valve is faulty (short circuit)
Remedy
Contact us.
Fault
Warning 193.01 - Valve Missing
Cause
The MB Stator cooling solenoid valve is disconnected or missing (open circuit)
Remedy
Make sure that the DP motor cooling solenoid valve is installed and is correctly
connected.
Fault
Warning 193.01 - Valve Fault
Cause
The MB Stator cooling solenoid valve is faulty (short circuit)
Remedy
Contact us.
10/2020 - ©Edwards Limited
Page 75
M56635880_D
M56635880_D - Fault finding
Содержание iXM Series
Страница 1: ...Dry Pump System iXM INSTRUCTION MANUAL M56635880_D Original instructions edwardsvacuum com...
Страница 101: ...This page has been intentionally left blank...
Страница 102: ...This page has been intentionally left blank...
Страница 103: ...This page has been intentionally left blank...
Страница 104: ...edwardsvacuum com...