![Edwards iXM Series Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/edwards/ixm-series/ixm-series_instruction-manual_2366129046.webp)
Figure 16 Protective Earth (PE) connection
Electrical connector locking mechanism
This product is approved permanently connected equipment and must be used as
described below:
The pumping system has an electrical connector locking mechanism (
connectors on the rear of the system
). Use an applicable screwdriver to release.
The pumping system is supplied with a protective cover (
) installed and the locking mechanism can be applied. Use the
following instructions to install the electrical supply cable.
▪
Turn the locking screws to loosen the locking mechanism, if necessary.
▪
Move the locking mechanism as far as possible to the left.
▪
Push up the lever to release the protective cover.
▪
Remove the protective cover and attach the electrical supply cable.
▪
Pull the lever downwards to lock the electrical supply cable connector in place.
▪
Push the locking mechanism as far as possible to the right so that it prevents the
lever from functioning.
▪
Tighten the locking screws to firmly hold the locking mechanism in place.
▪
Connect the other end of the electrical supply cable to the electrical supply
through an isolator.
10/2020 - ©Edwards Limited
Page 43
M56635880_D
M56635880_D - Installation
Содержание iXM Series
Страница 1: ...Dry Pump System iXM INSTRUCTION MANUAL M56635880_D Original instructions edwardsvacuum com...
Страница 101: ...This page has been intentionally left blank...
Страница 102: ...This page has been intentionally left blank...
Страница 103: ...This page has been intentionally left blank...
Страница 104: ...edwardsvacuum com...