Figure 17 Electrical connector locking mechanism (iXM200, iXM600 and iXM1200)
1. Electrical connector locking mechanism
2. Locking screws
3. Lever
4. Protective cover
1. Electrical connector locking mechanism
2. Locking screws
3. Lever
4. Protective cover
Figure 18 Electrical connector locking mechanism
1. Electrical connector locking mechanism
2. Pivot screw
3. Locking screw
4. Lever (2 off)
5. Protective cover
1. Electrical connector locking mechanism
2. Pivot screw
3. Locking screw
4. Lever (2 off)
5. Protective cover
4.6.7 Connect an additional RF earth (ground) (optional)
We recommend that you to connect an additional RF (Radio Frequency) earth if the
system is operated in an area subject to high RF emissions:
1.
Use a star washer to connect the end of the earth (ground) cable (
) connected to the dry pumping system inlet to one of the bolts
used to secure the inlet flange.
2.
Connect an additional earth (ground) cable to the RF earth (ground) stud (
Controls/connectors on the rear of the system
). An applicable low impedance cable
must be used (for example, use braided cable).
10/2020 - ©Edwards Limited
Page 44
M56635880_D
M56635880_D - Installation
Содержание iXM Series
Страница 1: ...Dry Pump System iXM INSTRUCTION MANUAL M56635880_D Original instructions edwardsvacuum com...
Страница 101: ...This page has been intentionally left blank...
Страница 102: ...This page has been intentionally left blank...
Страница 103: ...This page has been intentionally left blank...
Страница 104: ...edwardsvacuum com...