DYSIS Ultra 2.0 Digital Colposcope Instructions for Use
0230-53095, Rev 05
Revision Date 05-Mar-2021
Page
76
of
79
31.
Appendix 2: 60601-1-2 DATA
Tables of guidance and manufacturer’s declaration regarding electromagnetic emissions and immunity as per EN
60601-1-2:2015. Data extracted from Eurofins York EMC Test Report 5109TC2.
EN60601-1-2: 2015
Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard:
Electromagnetic disturbances
–
Requirements and tests
For Professional Healthcare Facility Environment
Reference Standard
Levels
Result
Emissions
Conducted RF emission
EN55011:2009+ A1:2010
Referenced in Table 2 (Page 26)
Class A
Mode 1
Mode 2
Pass
Radiated RF emission (Enclosure)
EN55011:2009+ A1:2010
Referenced in Table 2 (Page 26)
Class A
Mode 1
Mode 2
Pass
Harmonic Distortion
EN61000-3-2+A1:2009+A2:2009
Referenced in Table 2 (Page 26) and section 7.2.1
A
Mode 1
Pass
Voltage Fluctuations and Flicker
EN61000-3-3: 2013
Referenced in Table 2 (Page 26) and section 7.2.2
Plt
Pst
Dmax
Mode 1
Pass
Immunity
Enclosure Port (Table 4
–
Page 34)
Electrostatic Discharge
EN61000-4-2: 2009
Referenced in Table 4 (Page 34)
2, 4, 8, 15kV air
2, 4, 6, 8kV contact
Mode 1
Pass
Radiated RF EM fields
EN61000-4-3: 2006+A1:2008+ IS1:2009+A2: 2010
Referenced in Table 4 (Page 34)
3V/m
80-2700MHz
Mode 1
Pass
Proximity fields from RF wireless communications equipment
EN61000-4-3: 2006+A1:2008+ IS1:2009+A2: 2010
Referenced in Table 9 (Page 40)
As detailed below and in section 8.10
of standard
Mode 1
Pass