DYSIS Ultra 2.0 Digital Colposcope Instructions for Use
0230-53095, Rev 05
Revision Date 05-Mar-2021
Page
54
of
79
Ensure there is no ingress of fluid into connection ports, switches and buttons. Cleaning of the monitor
touchscreen and optical head are dealt with separately
25.4.
Touchscreen Monitor Cleaning
The touchscreen monitor can be cleaned with commercially available disinfectant / antiviral cleaning wipes or
a damp cloth moistened with a mild detergent / antiviral solution and dried. Excessive moisture or pressure
should not be used.
25.5.
Optical Head Cleaning
The optical head can be cleaned by wiping with a damp cloth moistened with mild detergent / antiviral
solution.
The glass cover over the camera lens at the front of the optical head can be cleaned with a cotton cloth or lens
tissue moistened with isopropyl alcohol or lens cleaner. Pay special attention not to scratch the surface during
cleaning
IMPORTANT NOTE: DO NOT USE CIRCULAR MOTION WHEN WIPING THE WINDOW; USE ONLY STRAIGHT
MOVEMENTS
–
EITHER UP AND DOWN OR SIDE TO SIDE.
25.6.
Disinfection
If needed the DYSIS Ultra 2.0 Digital Colposcope can be wiped with a soft cloth dampened with 70%
isopropanol alcohol, disinfectant wipes (chlorhexidine for viral control) and similar substances. DYSIS Ultra 2.0
Digital Colposcope is not intended to be sterilized. Refer to cleaning instructions in 0230-53059 - DYSIS Ultra
2.0 Digital Colposcopes Cleaning and Maintenance Instructions.
25.7.
DYSIS Specula
DYSIS Specula are either reusable stainless steel or disposable single patient use devices. After use, disposable
specula should be disposed of according to local clinical guidelines. Reusable specula should be cleaned
according to document 0230-53004 DYSIS Reusable Care and Maintenance Instructions and 0230-53028 DYSIS
disposable speculum IFU.