DYSIS Ultra 2.0 Digital Colposcope Instructions for Use
0230-53095, Rev 05
Revision Date 05-Mar-2021
Page
57
of
79
Colposcope
’
s base may be
loose/disconnected
Computing unit
malfunction
Contact DYSIS Medical service
department
Computing unit turns on, the
software functions, but the
DYSIS Exam button is disabled
Camera malfunction
Shut down the device and restart. If
the problem persists, contact DYSIS
Medical service department
Dynamic Imaging cannot be
initiated
Error in plunger
Check that there is sufficient acetic
acid solution in the container. From
the
home menu, press “pu
rge
syringe”. If problem persists
, contact
DYSIS Medical service department.
System’s locale is setup as
US and calibration is
required.
Contact service to reconfigure
system to UK settings
Image is displayed, but
illumination does not turn on
Illumination source
malfunction
Contact DYSIS Medical service
department
DYSIS Ultra 2.0 Digital
Colposcope
doesn’t turn on
Main power cable is
disconnected
Ensure the power switch is OFF and
reconnect the main power cable to
the power brick & the power cable is
locked into the DYSIS Ultra 2.0 Digital
Colposcope base. Turn the main
power switch ON
Outlet has no electricity
Ensure the electrical outlet is
operational
System malfunction
Contact service support
Calibration cannot be
performed
The card is very dirty
Contact DYSIS Medical service
department
System malfunction
Contact DYSIS Medical service
department
Software update cannot be
performed
The USB stick is not
recognized
Use another USB stick of ex-FAT
format
The configured network
path in which the upgrader
is, is not accessible.
Check connections through the
connectivity tab of settings to ensure
the configured network path is
accessible.
The DYSIS Ultra 2.0 Digital
Colposcope
software
upgrade is not on the USB
stick
Follow software update instructions
carefully. The upgrade file should not
be embedded in a folder on the USB.
System malfunction
Contact DYSIS Medical service
department
Power cable cannot be
plugged in to DYSIS Ultra 2.0
Digital Colposcope
The inner circle is pressed,
and power cable cannot
lock on DYSIS Ultra 2.0
Digital Colposcope
Hold the outer circle and pull back
the inner circle